7T 3D ‐ EPI PCASL With High SNR Efficiency and Robustness to Through‐Plane B 0 Field Gradients
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70491 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Gael Saib und 2 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Functional and Molecular Imaging, National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, Mar…. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Echo-Planar Imaging, Brain. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NINDS NIH HHS, Intramural Research Program of the National Institutes of Health, National Institute of Neurological Disorders and Stroke.
Abstract (Originalsprache)
Purpose
7T pseudo‐continuous arterial spin labeling (PCASL) can benefit from combining 3D‐EPI with high‐performance background suppression (BS) pulses. However, the perfusion signal can be compromised by increased B 1 and B 0 inhomogeneity, and high SAR demands limiting the labeling duration (LD), labeling efficiency and BS inversion efficiency.
Methods
7T PCASL SNR efficiency (SNReff) was quantified using a range of LD (500–4000 ms) with optimized BS inversion pulses and 3D‐EPI acquisitions. PCASL sensitivity to B 0 was also investigated by measuring through‐plane field variations along the feeding arteries and acquiring perfusion data across a range of PCASL mean gradients ( G mean , −0.8 to +0.8 mT/m) with in‐plane B 0 corrections.
Results
Optimized WURST BS pulses achieved ∼0.93 inversion efficiency. PCASL SNReff was maximized when employing a LD of ∼3 s (B 1pcasl = 1 T, duty cycle = 60%), providing ∼25% higher SNReff than 1 s LD typically used at 7T. Arteries with strong negative though‐plane B 0 gradients () at the labeling plane generated low perfusion signal in the corresponding arterial territories when using a G mean of 0.4 mT/m. In these cases, the perfusion signal increased ∼128% by using a higher G mean of 0.8 mT/m.
Conclusion
7T PCASL SNReff can be maximized with LD of ∼3 s when using 3D‐EPI and optimized WURST BS pulses. A G mean of 0.8 mT/m improved PCASL robustness to B 0 field gradients along the feeding arteries without the need for subject specific corrections. Use of higher G mean also offers benefits in cases with multiple polarities and non‐linear field variations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Laboratory of Functional and Molecular Imaging, National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, Maryland, USA.
- NIH MRI Research Facility, National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, Maryland, USA.
ORCID-Profile: Gael Saib
Förderung
- NINDS NIH HHS
- Intramural Research Program of the National Institutes of Health
- National Institute of Neurological Disorders and Stroke
- National Institutes of Health
Zitieren
APA
Saib, G., Koretsky, A. P., & Talagala, S. L. (2026). 7T 3D ‐ EPI PCASL With High SNR Efficiency and Robustness to Through‐Plane B 0 Field Gradients. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70491
Vancouver
Saib G, Koretsky AP, Talagala SL. 7T 3D ‐ EPI PCASL With High SNR Efficiency and Robustness to Through‐Plane B 0 Field Gradients. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70491
BibTeX
@article{Saib2026,
author = {Saib, Gael and Koretsky, Alan P and Talagala, S Lalith},
title = {7T 3D ‐ EPI PCASL With High SNR Efficiency and Robustness to Through‐Plane B 0 Field Gradients},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70491}
}
EUVIMED