EUVIMEDEUVIMED entdecken
Open Access

Depth effect on point shear wave velocity elastography: Evidence in a chronic hepatitis C patient cohort

Leonardo Rizzo, Luca L'Abbate, Massimo Attanasio iD, Arturo Montineri, Salvatore Magliocco, Vincenza Calvaruso

Ultrasound · 2024-02 · 32 (1) · 53-61

DOI: 10.1177/1742271x231183370 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Ultrasound, online erschienen am 15.09.2023. Verfasst von Leonardo Rizzo und 5 weiteren Autor:innen. Beteiligte Einrichtung: Ultrasuoni Srl, Medical Center, Catania, Italy. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Aims: This study investigated the depth-related bias and the influence of scan plane angle on performance of point-shear-wave elastometry in a chronic hepatitis C patient cohort. Materials and Methods: We included 104 patients affected by chronic liver disease related to the hepatitis C virus. Liver surface nodularity was the reference to diagnose cirrhosis. The ultrasound platform was the Siemens S2000, equipped with point-shear-wave elastometry software. Measurements were obtained in left lateral decubitus from the liver surface to the maximum depth of 8 cm in two orthogonal scan planes according to a standard sampling plane. Scatterplot and box plots explored the depth-related bias graphically. The area under the receiver operating characteristic was used to determine the point-shear-wave elastometry diagnostic performance at progressive depths according to liver surface nodularity.

Results

Of the 104 patients, 68 were cirrhotics. Depth-related bias equally modified point-shear-wave elastometry in the two orthogonal scan planes. A better point-shear-wave elastometry diagnostic performance was observed between depths of 4 and 5 cm. The frontal scan plane assured better discrimination between cirrhotic patients and non-cirrhotic patients.

Conclusion

Depth is crucial for point-shear-wave elastometry performance. Excellent diagnostic performance at a depth between 4 and 5 cm can also be obtained with a smaller number of measurements than previously recommended.

Schlagwörter

Stichwörter der Autor:innen

C hepatitisCirrhosisdepthliver fibrosisliver surface nodularity

Autor:innen und Institutionen

  • Ultrasuoni Srl, Medical Center, Catania, Italy.
  • Department of Biomedical and Dental Sciences and Morphofunctional Imaging, University of Messina, Messina, Italy.
  • Department of Economics, Business and Statistics, University of Palermo, Palermo, Italy.
  • Department of Infectious Diseases and Tropical Medicine, AOU Policlinico 'G. Rodolico - San Marco', Catania, Italy.
  • Gastroenterology and Hepatology Unit, Department of Health Promotion Sciences Maternal and Infantile Care, Internal Medicine and Medical Specialities, PROMISE, University of Palermo, Palermo, Italy.

ORCID-Profile: Massimo Attanasio

Zitieren

APA

Rizzo, L., L'Abbate, L., Attanasio, M., Montineri, A., Magliocco, S., & Calvaruso, V. (2024). Depth effect on point shear wave velocity elastography: Evidence in a chronic hepatitis C patient cohort. Ultrasound, 32(1), 53-61. https://doi.org/10.1177/1742271x231183370

Vancouver

Rizzo L, L'Abbate L, Attanasio M, Montineri A, Magliocco S, Calvaruso V. Depth effect on point shear wave velocity elastography: Evidence in a chronic hepatitis C patient cohort. Ultrasound. 2024;32(1):53-61. doi:10.1177/1742271x231183370

BibTeX
@article{Rizzo2024,
  author = {Rizzo, Leonardo and L'Abbate, Luca and Attanasio, Massimo and Montineri, Arturo and Magliocco, Salvatore and Calvaruso, Vincenza},
  title = {Depth effect on point shear wave velocity elastography: Evidence in a chronic hepatitis C patient cohort},
  journal = {Ultrasound},
  year = {2024},
  volume = {32},
  number = {1},
  pages = {53--61},
  doi = {10.1177/1742271x231183370}
}
Kostenlos starten →