Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 11.07.2024. Verfasst von Gianpaolo Vidili und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medicine, Surgery and Pharmacy, University of Sassari, 07100 Sassari, Italy. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Transient Elastography (TE) is widely regarded as the most reliable non-invasive method for evaluating liver fibrosis. Recently, new techniques such as 2D Shear Wave Elastography (2D-SWE) have been developed. This study aimed to evaluate the correlation between TE and 2D-SWE in patients with HCV-related chronic liver disease and to redefine the cut-off values of 2D-SWE for predicting different stages of fibrosis based on our results.
Methods
Both TE (Fibroscan, Echosens, Paris, France) and 2D-SWE (SuperSonic Imagine) were performed simultaneously in 170 patients, including those with active and eradicated HCV infection. Spearman’s rank correlation coefficient was used to assess the correlation between the two measurements, and the concordance between the assigned METAVIR classes was calculated using Cohen’s kappa coefficient. ROC curves were constructed to determine the optimal cut-off values for 2D-SWE.
Results
Ten patients were excluded for invalid measurements. In the remaining 160 patients, TE and 2D-SWE demonstrated a high correlation (ρ = 0.83, p < 0.0001) and good agreement in METAVIR classification (k = 0.74). The optimal cut-off values identified for 2D-SWE were as follows: ≥ 7 kPa for F ≥ 2, ≥ 8.3 kPa for F ≥ 3, and ≥ 9.4 kPa for F4.
Conclusions
2D-SWE is a viable alternative to TE for patients with HCV-related chronic liver disease. Our data suggest that the currently accepted 2D-SWE cut-off values for cirrhosis (F4) should be reconsidered and potentially lowered.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medicine, Surgery and Pharmacy, University of Sassari, 07100 Sassari, Italy.
- Department of Internal Medicine, Azienda Ospedaliero Universitaria di Sassari, 07100 Sassari, Italy.
- Centralized Day Hospital of the Medical Area, Azienda Ospedaliero Universitaria di Sassari, viale San Pietro 8, 07100 Sassari, Italy.
- Department of Biomedical Sciences, Public Health-Laboratory of Biostatistics, University of Sassari, 07100 Sassari, Italy.
ORCID-Profile: Marco Arru, Pierluigi Meloni, Giuliana Solinas, Sebastiana Atzori, Ivana Maida
Zitieren
APA
Vidili, G., Arru, M., Meloni, P., Solinas, G., Atzori, S., & Maida, I. (2024). Comparison of 2D Shear Wave Elastography and Transient Elastography in Non-Invasive Evaluation of Liver Fibrosis in Hepatitis C Virus-Related Chronic Liver Disease. Journal of Clinical Medicine, 13(14), 4061. https://doi.org/10.3390/jcm13144061
Vancouver
Vidili G, Arru M, Meloni P, Solinas G, Atzori S, Maida I. Comparison of 2D Shear Wave Elastography and Transient Elastography in Non-Invasive Evaluation of Liver Fibrosis in Hepatitis C Virus-Related Chronic Liver Disease. Journal of Clinical Medicine. 2024;13(14):4061. doi:10.3390/jcm13144061
BibTeX
@article{Vidili2024,
author = {Vidili, Gianpaolo and Arru, Marco and Meloni, Pierluigi and Solinas, Giuliana and Atzori, Sebastiana and Maida, Ivana},
title = {Comparison of 2D Shear Wave Elastography and Transient Elastography in Non-Invasive Evaluation of Liver Fibrosis in Hepatitis C Virus-Related Chronic Liver Disease},
journal = {Journal of Clinical Medicine},
year = {2024},
volume = {13},
number = {14},
pages = {4061},
doi = {10.3390/jcm13144061}
}
EUVIMED