EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Mapping tendon stiffness: Two-dimensional versus point shear wave elastography

Mario Pasurka, Theo Falck, Joshua Kubach, Stefan Söllner, Deike Strobel, Mario Perl, Marcel Betsch

Journal of Orthopaedics · 2025-12 · 70 · 276-282

DOI: 10.1016/j.jor.2025.08.027 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Orthopaedics, 2025-12. Verfasst von Mario Pasurka und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic and Trauma Surgery, University Hospital of Erlangen, Friedrich-Alexander-University Erlangen-Nuremberg, Krankenhaus 9, 9105…. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Rationale and objectives: To date, no study has investigated the comparability between point shear wave elastography (pSWE) and 2D-shear wave elastography (2D-SWE) in assessing patellar tendon (PT) stiffness. In addition to comparing shear wave velocities, this study also evaluated intra- and interrater reliability, as well as measurement time. Materials and methods: ) were recruited as participants. Each participant underwent a standardized multimodal ultrasound protocol that included B-mode ultrasound (B-US), Color Doppler ultrasound (CD-US), and a shear wave elastography (SWE) examination of the bilateral patellar tendon using both pSWE and 2D-SWE. Stiffness values and examination times were recorded. Intra- and interrater reliability were assessed across two different measurement time points with two investigators for both pSWE and 2D-SWE.

Results

Elastography values were slightly higher for pSWE compared to 2D-SWE (pSWE: 4.02 ± 1.07 m/s, 2D-SWE: 3.82 ± 0.73 m/s; p < 0.001). 2D-SWE10 measurements took significantly more time than pSWE (2D-SWE10: 69.89 ± 13.46 s vs. pSWE: 54.85 ± 8.34 s, p < 0.001). However, the time required for 2D-SWE1 was significantly shorter than for pSWE (2D-SWE1: 47.20 ± 8.23 s vs. pSWE: 54.85 ± 8.34 s, p < 0.001). Both interrater and intrarater reliability reached excellent levels for all SWE measurements, with the highest interrater and intrarater ICC values obtained for 2D-SWE1.

Conclusion

This is the first study to evaluate and compare pSWE and 2D-SWE in the assessment of patellar tendon stiffness. These findings could have important implications for the clinical use of musculoskeletal SWE, providing insights into measurement time, reliability, and the potential benefits of using 2D-SWE in clinical practice.

Schlagwörter

Stichwörter der Autor:innen

Acoustic radiation force impulse imagingMusculoskeletal ultrasoundPoint shear-wave elastographyTendon stiffnessTwo-dimensional shear wave elastography

Autor:innen und Institutionen

  • Department of Orthopaedic and Trauma Surgery, University Hospital of Erlangen, Friedrich-Alexander-University Erlangen-Nuremberg, Krankenhaus 9, 91054, Erlangen, Germany.
  • Department of Medicine 1, University Hospital Erlangen, Friedrich-Alexander-University Erlangen-Nuremberg, Ulmenweg 18, 91054, Erlangen, Germany.

Zitieren

APA

Pasurka, M., Falck, T., Kubach, J., Söllner, S., Strobel, D., Perl, M., & Betsch, M. (2025). Mapping tendon stiffness: Two-dimensional versus point shear wave elastography. Journal of Orthopaedics, 70, 276-282. https://doi.org/10.1016/j.jor.2025.08.027

Vancouver

Pasurka M, Falck T, Kubach J, Söllner S, Strobel D, Perl M, et al. Mapping tendon stiffness: Two-dimensional versus point shear wave elastography. Journal of Orthopaedics. 2025;70:276-282. doi:10.1016/j.jor.2025.08.027

BibTeX
@article{Pasurka2025,
  author = {Pasurka, Mario and Falck, Theo and Kubach, Joshua and Söllner, Stefan and Strobel, Deike and Perl, Mario and Betsch, Marcel},
  title = {Mapping tendon stiffness: Two-dimensional versus point shear wave elastography},
  journal = {Journal of Orthopaedics},
  year = {2025},
  volume = {70},
  pages = {276--282},
  doi = {10.1016/j.jor.2025.08.027}
}
Kostenlos starten →