The impact of arterial catheter-to-vessel ratio on distal arterial flow and measured arterial pressures in a pulsatile arterial vessel phantom
The Journal of Vascular Access · 2025
DOI: 10.1177/11297298251383359 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Vascular Access, 2025. Verfasst von Mark D Weber und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology and Critical Care Medicine, Children's Hospital of Philadelphia, Philadelphia, PA, USA. Hauptthemen laut MEDLINE: Pulsatile Flow, Arterial Pressure, Catheterization, Peripheral, Vascular Access Devices. Die Arbeit nennt 20 Literaturangaben. Gefördert von: National Institutes of Health.
Abstract (Originalsprache)
Background
The effect of catheter size on distal arterial flow and measured arterial pressure remains underexplored. This study evaluates the relationship between catheter-to-vessel ratio (CVR), distal arterial flow, and measured blood pressures in a controlled pulsatile arterial vessel phantom.
Methods
Using in vitro data simulating arterial conditions, distal peak systolic velocity, systolic (SBP), and diastolic (DBP) were analyzed across varying catheter sizes and vessel diameters. Pearson correlation and linear regression were used to assess the relationship between CVR and changes in relative distal velocity, SBP, and DBP.
Results
Relative distal peak systolic velocity exhibited a strong inverse relationship with CVR ( r = −0.903, p < 0.001), with linear regression indicating that for every 10% increase in CVR, distal peak systolic velocity decreased by approximately 9.3%. Flow reduced >20% when CVR exceeded 0.6 and 10 mmHg when CVR exceeded 0.6 and
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anesthesiology and Critical Care Medicine, Children's Hospital of Philadelphia, Philadelphia, PA, USA.
- Department of Advanced Clinical Practice, Children's Hospital of Philadelphia, Philadelphia, PA, USA.
- Department of Anesthesiology and Critical Care Medicine, Perelman School of Medicine at the University of Pennsylvania, Children's Hospital of Philadelphia, Philadelphia, PA, USA.
- Central Bucks School District, Doylestown, PA, USA.
ORCID-Profile: Mark D Weber
Förderung
- National Institutes of Health
Zitieren
APA
Weber, M. D., Senthil, K., Zuckerberg, J. C., Weber, S. T., Himebauch, A. S., & Conlon, T. W. (2025). The impact of arterial catheter-to-vessel ratio on distal arterial flow and measured arterial pressures in a pulsatile arterial vessel phantom. The Journal of Vascular Access. https://doi.org/10.1177/11297298251383359
Vancouver
Weber MD, Senthil K, Zuckerberg JC, Weber ST, Himebauch AS, Conlon TW. The impact of arterial catheter-to-vessel ratio on distal arterial flow and measured arterial pressures in a pulsatile arterial vessel phantom. The Journal of Vascular Access. 2025. doi:10.1177/11297298251383359
BibTeX
@article{Weber2025,
author = {Weber, Mark D and Senthil, Kumaran and Zuckerberg, Jeremy C and Weber, Silas T and Himebauch, Adam S and Conlon, Thomas W},
title = {The impact of arterial catheter-to-vessel ratio on distal arterial flow and measured arterial pressures in a pulsatile arterial vessel phantom},
journal = {The Journal of Vascular Access},
year = {2025},
doi = {10.1177/11297298251383359}
}
EUVIMED