EUVIMEDEUVIMED entdecken
Crossref

Visualizing Flow Dynamics in a Human Aorta During Peripheral Veno‐Arterial Extracorporeal Membrane Oxygenation Treatment Through Particle Image Velocimetry Study

Dhayananth Kanagarajan iD, Clayton Semenzin, Silver Heinsar iD, Van Thanh Dau, Jo P Pauls iD, Geoffrey D Tansley, John F Fraser

Artificial Organs · 2025

DOI: 10.1111/aor.15040 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Artificial Organs, 2025. Verfasst von Dhayananth Kanagarajan und 6 weiteren Autor:innen. Beteiligte Einrichtung: School of Engineering and Built Environment, Griffith University, Gold Coast, Queensland, Australia. Hauptthemen laut MEDLINE: Extracorporeal Membrane Oxygenation, Aorta, Models, Cardiovascular. Die Arbeit nennt 21 Literaturangaben. Gefördert von: Griffith University, Estonian Research Council, Prince Charles Hospital Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To build a mock circulatory loop to simulate veno‐arterial extracorporeal membrane oxygenation (V‐A ECMO) haemodynamics using a patient‐specific silicone aortic phantom and compare the flow dynamics for continuous and pulsatile ECMO settings.

Methods

A silicone model of a human aorta with major branches was fabricated and the left iliac artery was cannulated using a Xenios ECMO device to simulate a peripheral V‐A ECMO configuration. The left ventricular outflows at the aortic root were simulated using a piston pump. Particle image velocimetry (PIV) experiments were conducted to measure flow velocities along the aortic phantom for a severe heart failure condition (LV = 1 L/min) supported by ECMO (4 L/min) at a heart rate of 80 bpm.

Results

Comparing all ECMO support modes, counter pulse ECMO mode exhibited higher velocities during both peak systolic and mid‐diastolic phases. This shows that the LV and ECMO device supplied fluid into the aorta with less flow resistance, whereas other modes exhibited higher flow resistance due to the flow collision of antegrade LV and retrograde ECMO flow in the aorta.

Conclusion

Pulsatile flow ECMO with better placement of ECMO pulse delivery in the cardiac cycle could develop greater pulsatility into the aorta, which is crucial for transferring haemodynamic energy and reducing cardiac afterload.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Extracorporeal Membrane OxygenationAortaModels, CardiovascularHumansHemodynamicsPulsatile FlowRheologyBlood Flow VelocityHeart FailurePhantoms, Imaging

Stichwörter der Autor:innen

PIVV‐A ECMOextracorporeal membrane oxygenationmock circulatory looppulsatile flow

Autor:innen und Institutionen

  • School of Engineering and Built Environment, Griffith University, Gold Coast, Queensland, Australia.
  • Innovative Cardiovascular Engineering and Technology Laboratory, Critical Care Research Group, The Prince Charles Hospital, Brisbane, Queensland, Australia.
  • Department of Intensive Care, North Estonia Medical Centre, Tallinn, Estonia.
  • Department of Cardiology, Institute of Clinical Medicine, University of Tartu, Tartu, Estonia.
  • Faculty of Medicine, University of Queensland, Brisbane, Queensland, Australia.
  • School of Medicine, Griffith University, Gold Coast, Queensland, Australia.

ORCID-Profile: Dhayananth Kanagarajan, Silver Heinsar, Jo P Pauls

Förderung

  • Griffith University
  • Estonian Research Council
  • Prince Charles Hospital Foundation
  • Eesti Teadusagentuur
  • Common Good

Zitieren

APA

Kanagarajan, D., Semenzin, C., Heinsar, S., Dau, V. T., Pauls, J. P., Tansley, G. D., & Fraser, J. F. (2025). Visualizing Flow Dynamics in a Human Aorta During Peripheral Veno‐Arterial Extracorporeal Membrane Oxygenation Treatment Through Particle Image Velocimetry Study. Artificial Organs. https://doi.org/10.1111/aor.15040

Vancouver

Kanagarajan D, Semenzin C, Heinsar S, Dau VT, Pauls JP, Tansley GD, et al. Visualizing Flow Dynamics in a Human Aorta During Peripheral Veno‐Arterial Extracorporeal Membrane Oxygenation Treatment Through Particle Image Velocimetry Study. Artificial Organs. 2025. doi:10.1111/aor.15040

BibTeX
@article{Kanagarajan2025,
  author = {Kanagarajan, Dhayananth and Semenzin, Clayton and Heinsar, Silver and Dau, Van Thanh and Pauls, Jo P and Tansley, Geoffrey D and Fraser, John F},
  title = {Visualizing Flow Dynamics in a Human Aorta During Peripheral Veno‐Arterial Extracorporeal Membrane Oxygenation Treatment Through Particle Image Velocimetry Study},
  journal = {Artificial Organs},
  year = {2025},
  doi = {10.1111/aor.15040}
}
Kostenlos starten →