Frame-wise multi-echo distortion correction for superior functional MRI

Andrew N Van iD, David F Montez, Timothy O Laumann iD, Philip N Cho iD, Vahdeta Suljic, Thomas Madison iD, Noah J Baden iD, Nadeshka Ramirez-Perez, Kristen M Scheidter, Julia S Monk, Forrest I Whiting, Babatunde Adeyemo und 21 weitere

Imaging Neuroscience · 29.05.2026 · 4

DOI: 10.1162/imag.a.1262 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Imaging Neuroscience, online erschienen am 29.05.2026. Verfasst von Andrew N Van und 32 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Washington University in St. Louis, St. Louis, MO, United States. Die Arbeit nennt 62 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIMH NIH HHS, National Institute of Neurological Disorders and Stroke, National Institute of Mental Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Functional MRI (fMRI) data are severely distorted by magnetic field (B0) inhomogeneities, which currently must be corrected using separately acquired field map data. However, changes in the head position of a participant across fMRI frames cause changes in the B0 field, preventing accurate correction of geometric distortions. Movement during field map acquisitions corrupts field maps, preventing distortion correction altogether. In this study, we use multi-echo (ME) fMRI data to dynamically sample and correct for magnetic field image distortions caused by head motion. Our distortion correction pipeline, MEDIC (Multi-Echo DIstortion Correction), leverages magnetic field inhomogeneity information found in the difference between echoes and uses it to correct for distortion on a frame-by-frame basis. Here, we demonstrate that MEDIC’s frame-wise distortion correction decreases the impact of head motion on resting-state functional connectivity (RSFC) maps and improves alignment to anatomy when compared with the prior gold standard approach (i.e., FSL TOPUP). Enhanced frame-wise distortion correction with MEDIC, without the requirement for field map collection, furthers the benefit of cutting-edge multi-echo fMRI imaging over single-echo fMRI.

Schlagwörter

Stichwörter der Autor:innen

distortion correctionfMRImulti-echo

Autor:innen und Institutionen

  • Department of Biomedical Engineering, Washington University in St. Louis, St. Louis, MO, United States.
  • Department of Neurology, Washington University School of Medicine, St. Louis, MO, United States.
  • Department of Psychiatry, Washington University School of Medicine, St. Louis, MO, United States.
  • Institute of Child Development, University of Minnesota Medical School, Minneapolis, MN, United States.
  • Masonic Institute for the Developing Brain, University of Minnesota Medical School, Minneapolis, MN, United States.
  • Department of Radiology, Washington University School of Medicine, St. Louis, MO, United States.

ORCID-Profile: Andrew N Van, Timothy O Laumann, Philip N Cho, Thomas Madison, Noah J Baden, Roselyne J Chauvin, Athanasia Metoki, Aishwarya Rajesh, Jarod L Roland, Taylor Salo, Kimberly B Weldon, Aristeidis Sotiras, Joshua S Shimony, Benjamin P Kay, Brenden Tervo-Clemmens, Scott A Marek, Theodore D Satterthwaite, Evan M Gordon, Damien A Fair, M Dylan Tisdall, Nico U F Dosenbach

Förderung

  • NIMH NIH HHS
  • National Institute of Neurological Disorders and Stroke
  • National Institute of Mental Health
  • National Institute on Alcohol Abuse and Alcoholism
  • NIH

Zitieren

APA

Van, A. N., Montez, D. F., Laumann, T. O., Cho, P. N., Suljic, V., Madison, T., Baden, N. J., Ramirez-Perez, N., Scheidter, K. M., Monk, J. S., Whiting, F. I., Adeyemo, B., Chauvin, R. J., Krimmel, S. R., Metoki, A., Rajesh, A., Roland, J. L., Salo, T., Wang, A., …, & Dosenbach, N. U. F. (2026). Frame-wise multi-echo distortion correction for superior functional MRI. Imaging Neuroscience, 4. https://doi.org/10.1162/imag.a.1262

Vancouver

Van AN, Montez DF, Laumann TO, Cho PN, Suljic V, Madison T, et al. Frame-wise multi-echo distortion correction for superior functional MRI. Imaging Neuroscience. 2026;4. doi:10.1162/imag.a.1262

BibTeX
@article{Van2026,
  author = {Van, Andrew N and Montez, David F and Laumann, Timothy O and Cho, Philip N and Suljic, Vahdeta and Madison, Thomas and Baden, Noah J and Ramirez-Perez, Nadeshka and Scheidter, Kristen M and Monk, Julia S and Whiting, Forrest I and Adeyemo, Babatunde and Chauvin, Roselyne J and Krimmel, Samuel R and Metoki, Athanasia and Rajesh, Aishwarya and Roland, Jarod L and Salo, Taylor and Wang, Anxu and Weldon, Kimberly B and Sotiras, Aristeidis and Shimony, Joshua S and Kay, Benjamin P and Nelson, Steven M and Tervo-Clemmens, Brenden and Marek, Scott A and Vizioli, Luca and Yacoub, Essa and Satterthwaite, Theodore D and Gordon, Evan M and Fair, Damien A and Tisdall, M Dylan and Dosenbach, Nico U F},
  title = {Frame-wise multi-echo distortion correction for superior functional MRI},
  journal = {Imaging Neuroscience},
  year = {2026},
  volume = {4},
  doi = {10.1162/imag.a.1262}
}
Kostenlos starten →