Principles guiding PFA cardiac catheter and sheath management from preclinical assessment of microbubble formation and air intrusion models
Heart Rhythm · 2026-10 · 23 (10) · 2123-2131
DOI: 10.1016/j.hrthm.2026.05.016 ↗Kurz gefasst
Fachartikel in der Zeitschrift Heart Rhythm, 2026-10. Verfasst von Xiao-Dong Zhang und 20 weiteren Autor:innen. Beteiligte Einrichtung: Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center, Bronx, New York. Hauptthemen laut MEDLINE: Microbubbles, Embolism, Air, Atrial Fibrillation, Catheter Ablation. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Biosense Webster Inc, Johnson and Johnson.
Abstract (Originalsprache)
BACKGROUND: Pulsed field ablation (PFA) may create gaseous microemboli, with large-tip catheters being a potential risk factor for further air intrusion. However, scant characterization exists for how optimization of catheter and sheath handling affects the microembolic load when handling large-tip devices. Given that most PFA catheters for treatment of atrial fibrillation have large tips, strategies to reduce gaseous emboli are critical. OBJECTIVE: This study aimed to quantify microbubble formation and air intrusion of a PFA variable-loop circular catheter with a compatible guiding sheath during electrophysiology procedural steps under varying parametric conditions. METHODS: 3 preclinical models were used. In an in vivo porcine model, blood was redirected from the carotid artery into an extracorporeal loop connected to an ultrasonic microbubble detector. Similarly, a benchtop in vitro circulating model was used with a microbubble detector. Finally, negative pressures were simulated on the benchtop to evaluate sheath-handling steps and the potential for air intrusion as measured by a syringe. RESULTS: Catheter insertion through a sheath has a 7-fold higher proportion of bubbles >7 nL than ablation with the variable-loop circular catheter. Microbubble volume during insertion was reduced by 22%, 67%, and 78% with sheath submersion, irrigation, and aspiration, respectively. Air intrusion decreased by 55%, 56%, 65%, and 57% with sheath stopcock management, aspiration, irrigation, and slower sheath advancement, respectively. CONCLUSION: Among the steps evaluated, catheter advancement through the sheath constitutes the most substantial procedural factor that may contribute to gaseous embolization. Careful management of these devices may mitigate embolic risk.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center, Bronx, New York.
- Johnson & Johnson MedTech, Irvine, California.
- Department of Clinical Electrophysiology and Cardiac Pacing, Monzino Cardiological Center, IRCCS, Milan, Italy.
- Texas Cardiac Arrhythmia Institute, St David's Medical Center, Austin, Texas.
- Kansas City Heart Rhythm Institute, Kansas.
- Montefiore-Einstein Center for Heart and Vascular Care, Montefiore Medical Center, Bronx, New York. Electronic address: dibbia@gmail.com.
Förderung
- Biosense Webster Inc
- Johnson and Johnson
Zitieren
APA
Zhang, X. D., Zou, F., Gomez, T., Byun, E., Chen, Q., Marazzato, J., Schiavone, M., Mohanty, S., La Fazia, V. M., Motta, J., Rosales, C. Z., Pandey, S., Safren, L., Shafir, Y., Grupposo, V., Ynoa, D., Lin, A., Lakkireddy, D., Natale, A., …, & Di Biase, L. (2026). Principles guiding PFA cardiac catheter and sheath management from preclinical assessment of microbubble formation and air intrusion models. Heart Rhythm, 23(10), 2123-2131. https://doi.org/10.1016/j.hrthm.2026.05.016
Vancouver
Zhang XD, Zou F, Gomez T, Byun E, Chen Q, Marazzato J, et al. Principles guiding PFA cardiac catheter and sheath management from preclinical assessment of microbubble formation and air intrusion models. Heart Rhythm. 2026;23(10):2123-2131. doi:10.1016/j.hrthm.2026.05.016
BibTeX
@article{Zhang2026,
author = {Zhang, Xiao-Dong and Zou, Fengwei and Gomez, Tara and Byun, Eric and Chen, Qi and Marazzato, Jacopo and Schiavone, Marco and Mohanty, Sanghamitra and La Fazia, Vincenzo Mirco and Motta, Jishu and Rosales, Cristian Zamora and Pandey, Shubhi and Safren, Lowell and Shafir, Yuval and Grupposo, Vito and Ynoa, Domingo and Lin, Aung and Lakkireddy, Dhanunjaya and Natale, Andrea and Guttenplan, Nils and Di Biase, Luigi},
title = {Principles guiding PFA cardiac catheter and sheath management from preclinical assessment of microbubble formation and air intrusion models},
journal = {Heart Rhythm},
year = {2026},
volume = {23},
number = {10},
pages = {2123--2131},
doi = {10.1016/j.hrthm.2026.05.016}
}
EUVIMED