Influence of Prosthetic Interface and Manufacturing Metal on Mechanical Behavior of Two Extra Narrow Implants
The International Journal of Oral & Maxillofacial Implants · 2025
DOI: 10.11607/jomi.10960 ↗At a glance
Journal article in The International Journal of Oral & Maxillofacial Implants, 2025. Written by Eduardo Zancopé and 8 co-authors. Major MEDLINE topics: Dental Implants, Dental Prosthesis Design.
Abstract (original language)
PURPOSE: To evaluate the mechanical behavior of two types of dental implants (both 2.9 mm in diameter) using physical and virtual mechanical testing methodologies. MATERIALS AND METHODS: Two different extra-narrow Morse taper implant systems were evaluated: (1) Facility implants (2.9-mm diameter, 5-degree angulation of the internal conical portion) and (2) GM Narrow implants (2.9-mm diameter, 16-degree angulation of the internal conical portion). Both groups were evaluated on the results of the following methodologies: fractographic analyses and strength to failure, strain gauge, and thermomechanical cycling tests. RESULTS: Statistical analysis revealed a significant distinction (P < .001) between groups: The mechanical fracture efficiency of the narrow group was 25% greater than the Facility group. The examination using a scanning electron microscope revealed that all implants experienced fractures, with a tendency for these fractures to occur in areas with a lack of continuity at the implant-abutment interface. There was no difference between groups considering the strain around the cervical region of all implants tested (P = .987). The survival rate of crowns and prosthetic connections was 100%. There was no crown loosening, prosthetic component detachment, or fracture. CONCLUSIONS: The proposed modifications for the new implant and abutment system had a positive impact on the mechanical behavior of the implant, with a 25% increase in fracture resistance and better stress distribution. Clinically, the GM Narrow system has become more resistant, but both systems can securely be used in restricted areas with low masticatory effort, such as for lateral incisors and mandibular incisors.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Cite
APA
Zancopé, E., Zancopé, K., Pinto, A. V. C., Dib, C. N. T., Menegazzo, E. B., Naves, L. Z., Ribeiro, R. F., Felippi, C., & das Neves, F. D. (2025). Influence of Prosthetic Interface and Manufacturing Metal on Mechanical Behavior of Two Extra Narrow Implants. The International Journal of Oral & Maxillofacial Implants. https://doi.org/10.11607/jomi.10960
Vancouver
Zancopé E, Zancopé K, Pinto AVC, Dib CNT, Menegazzo EB, Naves LZ, et al. Influence of Prosthetic Interface and Manufacturing Metal on Mechanical Behavior of Two Extra Narrow Implants. The International Journal of Oral & Maxillofacial Implants. 2025. doi:10.11607/jomi.10960
BibTeX
@article{Zancop2025,
author = {Zancopé, Eduardo and Zancopé, Karla and Pinto, Ana Vitória Carvalho and Dib, Carolina Neves Tannous and Menegazzo, Eduarda Betiati and Naves, Lucas Zago and Ribeiro, Ricardo Faria and Felippi, Christian and das Neves, Flavio Domingues},
title = {Influence of Prosthetic Interface and Manufacturing Metal on Mechanical Behavior of Two Extra Narrow Implants},
journal = {The International Journal of Oral & Maxillofacial Implants},
year = {2025},
doi = {10.11607/jomi.10960}
}
EUVIMED