EUVIMEDEUVIMED entdecken
Fallbericht

Transcatheter Extraction and Implantation of a Novel Leadless Pacemaker via Tricuspid Valve‐in‐Valve: A Case Report

Zijia Wu, Shuyi Zeng, Zhihai Lin iD

Pacing and Clinical Electrophysiology · 14.06.2026

DOI: 10.1111/pace.70327 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Pacing and Clinical Electrophysiology, online erschienen am 14.06.2026. Verfasst von Zijia Wu und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Yulin First People's Hospital, The Sixth Affiliated Hospital of Guangxi Medical University, Yulin, China. Die Arbeit nennt 5 Literaturangaben. Gefördert von: Yulin City Scientific Research and Technological Development Plan.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Transcatheter retrieval of a dislodged leadless pacemaker (LP) from a tricuspid valve‐in‐valve (ViV) prosthesis has not been previously reported. Case Summary A 42‐year‐old woman with a prior bioprosthetic tricuspid valve, transvenous pacemaker, and subsequent percutaneous ViV implantation presented with ventricular lead dysfunction. A Micra LP was implanted but dislodged within 7 days. The device was retrieved using a sheath‐in‐sheath system combined with a double‐snare technique under fluoroscopic guidance. The first snare engaged the device, and traction induced tine deformation and device mobilization. After retraction into the right atrium, a second snare enabled coaxial alignment for safe retrieval into the introducer sheath. A new LP was immediately reimplanted with stable electrical parameters that remained satisfactory at 1‐year follow‐up. No prosthetic valve injury or thromboembolic complications occurred.

Conclusion

This case demonstrates that transcatheter retrieval and same‐day reimplantation of LP via a tricuspid ViV prosthesis is feasible and safe.

Schlagwörter

Stichwörter der Autor:innen

Leadless pacemakerdevice dislodgementdouble‐snare techniquetranscatheter retrievaltricuspid valve‐in‐valve

Autor:innen und Institutionen

  • Department of Cardiology, Yulin First People's Hospital, The Sixth Affiliated Hospital of Guangxi Medical University, Yulin, China.

ORCID-Profile: Zhihai Lin

Förderung

  • Yulin City Scientific Research and Technological Development Plan

Zitieren

APA

Wu, Z., Zeng, S., & Lin, Z. (2026). Transcatheter Extraction and Implantation of a Novel Leadless Pacemaker via Tricuspid Valve‐in‐Valve: A Case Report. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70327

Vancouver

Wu Z, Zeng S, Lin Z. Transcatheter Extraction and Implantation of a Novel Leadless Pacemaker via Tricuspid Valve‐in‐Valve: A Case Report. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70327

BibTeX
@article{Wu2026,
  author = {Wu, Zijia and Zeng, Shuyi and Lin, Zhihai},
  title = {Transcatheter Extraction and Implantation of a Novel Leadless Pacemaker via Tricuspid Valve‐in‐Valve: A Case Report},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2026},
  doi = {10.1111/pace.70327}
}
Kostenlos starten →