Transcatheter Extraction and Implantation of a Novel Leadless Pacemaker via Tricuspid Valve‐in‐Valve: A Case Report
Pacing and Clinical Electrophysiology · 2026-06-14
DOI: 10.1111/pace.70327 ↗At a glance
Fallbericht in Pacing and Clinical Electrophysiology, published online on 2026-06-14. Written by Zijia Wu and 2 co-authors. Institution: Department of Cardiology, Yulin First People's Hospital, The Sixth Affiliated Hospital of Guangxi Medical University, Yulin, China. 5 references. Funded by: Yulin City Scientific Research and Technological Development Plan.
Abstract (original language)
Background
Transcatheter retrieval of a dislodged leadless pacemaker (LP) from a tricuspid valve‐in‐valve (ViV) prosthesis has not been previously reported. Case Summary A 42‐year‐old woman with a prior bioprosthetic tricuspid valve, transvenous pacemaker, and subsequent percutaneous ViV implantation presented with ventricular lead dysfunction. A Micra LP was implanted but dislodged within 7 days. The device was retrieved using a sheath‐in‐sheath system combined with a double‐snare technique under fluoroscopic guidance. The first snare engaged the device, and traction induced tine deformation and device mobilization. After retraction into the right atrium, a second snare enabled coaxial alignment for safe retrieval into the introducer sheath. A new LP was immediately reimplanted with stable electrical parameters that remained satisfactory at 1‐year follow‐up. No prosthetic valve injury or thromboembolic complications occurred.
Conclusion
This case demonstrates that transcatheter retrieval and same‐day reimplantation of LP via a tricuspid ViV prosthesis is feasible and safe.
Subjects
Author keywords
Authors and institutions
- Department of Cardiology, Yulin First People's Hospital, The Sixth Affiliated Hospital of Guangxi Medical University, Yulin, China.
ORCID profiles: Zhihai Lin
Funding
- Yulin City Scientific Research and Technological Development Plan
Cite
APA
Wu, Z., Zeng, S., & Lin, Z. (2026). Transcatheter Extraction and Implantation of a Novel Leadless Pacemaker via Tricuspid Valve‐in‐Valve: A Case Report. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70327
Vancouver
Wu Z, Zeng S, Lin Z. Transcatheter Extraction and Implantation of a Novel Leadless Pacemaker via Tricuspid Valve‐in‐Valve: A Case Report. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70327
BibTeX
@article{Wu2026,
author = {Wu, Zijia and Zeng, Shuyi and Lin, Zhihai},
title = {Transcatheter Extraction and Implantation of a Novel Leadless Pacemaker via Tricuspid Valve‐in‐Valve: A Case Report},
journal = {Pacing and Clinical Electrophysiology},
year = {2026},
doi = {10.1111/pace.70327}
}
EUVIMED