Magnetic Resonance Imaging Safety in Patients With Conditional Mixed‐Manufacturer Devices
Pacing and Clinical Electrophysiology · 2026-11 · 49 (11) · 1736-1740
DOI: 10.1111/pace.70297 ↗At a glance
Journal article in Pacing and Clinical Electrophysiology, published online on 2026-05-22. Written by Ryka M Tashakkor and 5 co-authors. Institution: Southern California Heart Specialist, Arcadia, California, USA. Major MEDLINE topics: Magnetic Resonance Imaging, Pacemaker, Artificial, Defibrillators, Implantable, Patient Safety. 16 references. The full text is freely available (open access, CC BY-NC 4.0).
Abstract (original language)
Background
Magnetic resonance imaging (MRI) is a critical diagnostic tool known for its non‐invasive nature and high‐resolution imaging capabilities. However, its safety in patients with cardiac implantable electronic devices (CIEDs), particularly those with mixed‐lead systems, remains a significant clinical concern. This report investigates the safety and efficacy of MRI in such patients, where generators and leads originate from different manufacturers and are not officially deemed MRI‐conditional.
Methods
A retrospective, non‐randomized single center study was conducted involving 16 patients with mixed company lead CIEDs that obtained 23 MRIs. All patients underwent clinically indicated 1.5 Tesla MRI scans between July 2021 and August 2025.
Results
Pre‐ and post‐scan measurements included lead impedance, sensing thresholds, and battery voltage, along with monitoring for clinical complications such as arrhythmias, syncope, or cardiac arrest. No adverse effects or device malfunctions were observed, and all key parameters remained stable post‐scan.
Conclusion
This study shows the safe use of MRIs for patients with mixed‐lead CIEDs.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Southern California Heart Specialist, Arcadia, California, USA.
- Abbott, Sylmar, California, USA.
- University of Southern California, Arcadia, California, USA.
ORCID profiles: Ryka M Tashakkor, Kian Zamanian, Babak Tashakkor
Cite
APA
Tashakkor, R. M., Zamanian, K., Scott, N., Farhoumand, S., Hernandez, M., & Tashakkor, B. (2026). Magnetic Resonance Imaging Safety in Patients With Conditional Mixed‐Manufacturer Devices. Pacing and Clinical Electrophysiology, 49(11), 1736-1740. https://doi.org/10.1111/pace.70297
Vancouver
Tashakkor RM, Zamanian K, Scott N, Farhoumand S, Hernandez M, Tashakkor B. Magnetic Resonance Imaging Safety in Patients With Conditional Mixed‐Manufacturer Devices. Pacing and Clinical Electrophysiology. 2026;49(11):1736-1740. doi:10.1111/pace.70297
BibTeX
@article{Tashakkor2026,
author = {Tashakkor, Ryka M and Zamanian, Kian and Scott, Nathan and Farhoumand, Safa and Hernandez, Marisela and Tashakkor, Babak},
title = {Magnetic Resonance Imaging Safety in Patients With Conditional Mixed‐Manufacturer Devices},
journal = {Pacing and Clinical Electrophysiology},
year = {2026},
volume = {49},
number = {11},
pages = {1736--1740},
doi = {10.1111/pace.70297}
}
EUVIMED