Aortic velocity measurements derived from phase‐contrast MRI are influenced by a cardiac implantable electronic device in both adult and pediatric human subjects

Huili Yang iD, Oluyemi B Aboyewa iD, Gregory Webster, Dhaivat Shah iD, Laleh Golestanirad iD, Justin J Baraboo iD, Michael Markl, Jeremy D Collins, Bradley P Knight, KyungPyo Hong, Amit R Patel, Daniel C Lee und 1 weitere

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30399 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Huili Yang und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Northwestern University Feinberg School of Medicine, Chicago, Illinois, USA. Hauptthemen laut MEDLINE: Aorta, Defibrillators, Implantable, Pacemaker, Artificial, Artifacts. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIH HHS, NHLBI NIH HHS, American Heart Association.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Overall there is a lack of evidence on the accuracy and precision of phase‐contrast (PC) MRI in patients with cardiac implantable electronic devices (CIEDs). The purpose of this study is to determine whether aortic velocity measurements are influenced by a CIED.

Methods

We scanned 21 adult patients and 8 pediatric volunteers using clinical standard PC and real‐time PC (rt‐PC) sequences with and without a CIED generator taped on human subjects (below the left clavicle for adults and children; also on the abdomen for children) to mimic image artifacts. Peak and mean velocities above the aortic valve were calculated.

Results

The Bland–Altman analyses on peak velocity measurements in pediatric subjects showed that both the accuracy and precision worsen as the distance between the CIED and aortic valve decreases (i.e., from abdomen to below the left clavicle). Specifically, both the bias and the coefficient of variation (CV) for both clinical PC and rt‐PC increased from the abdominal position (clinical: bias = −1.1%, CV = 4.3%; rt‐PC: bias = −0.3%, CV = 3.4%) to the clavicle position (clinical: bias = −4.0%, CV = 8.1%; rt‐PC: bias = 8.2%, CV = 7.3%). A similar trend was observed for mean velocity. The mean difference in peak and mean velocity measurements between rt‐PC with CIED (either position) and clinical standard PC with no CIED was within 7.5%. In adult patients, the mean difference between rt‐PC with CIED and clinical standard PC with CIED in peak velocity was 6.9%, and the CV was 7.9%.

Conclusion

This study demonstrates that aortic velocity measurements are influenced by CIED in both adult and pediatric subjects.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AortaDefibrillators, ImplantablePacemaker, ArtificialArtifactsMagnetic Resonance ImagingHumansMaleAdultChildFemaleMiddle AgedReproducibility of ResultsBlood Flow VelocityAdolescentAgedSensitivity and SpecificityChild, PreschoolYoung Adult

Stichwörter der Autor:innen

CIEDaortic flowphase‐contrast MRIreal‐time MRI

Autor:innen und Institutionen

  • Department of Radiology, Northwestern University Feinberg School of Medicine, Chicago, Illinois, USA.
  • Department of Biomedical Engineering, McCormick School of Engineering, Northwestern University, Evanston, Illinois, USA.
  • Division of Cardiology, Department of Pediatrics, Ann & Robert H. Lurie Children's Hospital, Chicago, Illinois, USA.
  • Department of Radiology, Mayo Clinic, Rochester, 55905, Minnesota, USA.
  • Department of Medicine (Cardiology), Northwestern University Feinberg School of Medicine, Chicago, Illinois, USA.
  • Department of Medicine (Cardiovascular Division), University of Virginia School of Medicine, Charlottesville, Virginia, USA.

ORCID-Profile: Huili Yang, Oluyemi B Aboyewa, Dhaivat Shah, Laleh Golestanirad, Justin J Baraboo, Daniel Kim

Förderung

  • NIH HHS
  • NHLBI NIH HHS
  • American Heart Association
  • Radiological Society of North America
  • American Heart Association-American Stroke Association
  • National Institutes of Health

Zitieren

APA

Yang, H., Aboyewa, O. B., Webster, G., Shah, D., Golestanirad, L., Baraboo, J. J., Markl, M., Collins, J. D., Knight, B. P., Hong, K., Patel, A. R., Lee, D. C., & Kim, D. (2025). Aortic velocity measurements derived from phase‐contrast MRI are influenced by a cardiac implantable electronic device in both adult and pediatric human subjects. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30399

Vancouver

Yang H, Aboyewa OB, Webster G, Shah D, Golestanirad L, Baraboo JJ, et al. Aortic velocity measurements derived from phase‐contrast MRI are influenced by a cardiac implantable electronic device in both adult and pediatric human subjects. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30399

BibTeX
@article{Yang2025,
  author = {Yang, Huili and Aboyewa, Oluyemi B and Webster, Gregory and Shah, Dhaivat and Golestanirad, Laleh and Baraboo, Justin J and Markl, Michael and Collins, Jeremy D and Knight, Bradley P and Hong, KyungPyo and Patel, Amit R and Lee, Daniel C and Kim, Daniel},
  title = {Aortic velocity measurements derived from phase‐contrast MRI are influenced by a cardiac implantable electronic device in both adult and pediatric human subjects},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.30399}
}
Kostenlos starten →