Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2025. Verfasst von Arwa Younis und 19 weiteren Autor:innen. Beteiligte Einrichtung: Cardiac Electrophysiology Section, Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio, USA. Hauptthemen laut MEDLINE: Iatrogenic Disease, Device Removal, Pacemaker, Artificial, Heart Injuries. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 3 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The authors received no specific funding for this work..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Iatrogenic lead perforation is a rare but serious complication of cardiac implantable electronic device (CIED) implantation. Evidence on percutaneous management of subacute or delayed cases remains limited.

Methods

We retrospectively reviewed 38 patients treated for iatrogenic lead perforation between January 2012 and October 2024. Acute lead perforation cases, which were managed during the implantation procedure, were excluded. Lead removal was performed under imaging guidance in either the electrophysiology (EP) lab or a hybrid operating room (OR) with surgical backup. Clinical presentation, imaging, procedural approach, and outcomes were analyzed.

Results

Of 38 patients, 25 underwent lead removal in the EP lab and 13 in the OR. The mean age was 67 ± 15 years; 58% were female. Most perforated leads were pacemaker leads (58%), typically involving the right ventricle (72%). Complete removal and reimplantation were achieved in all patients using simple traction with or without a locking stylet. Pericardiocentesis was required in 8% of cases. No major complications or delayed events were observed over a median 16‐month follow‐up.

Conclusion

Percutaneous removal of perforated leads is feasible and safe in the EP lab when performed by experienced operators. Surgical backup should be readily available given the potential consequences of this complication.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Iatrogenic DiseaseDevice RemovalPacemaker, ArtificialHeart InjuriesDefibrillators, ImplantablePericardiocentesisHumansFemaleRetrospective StudiesAgedMaleTreatment OutcomeMiddle AgedTime FactorsAged, 80 and overRisk Factors

Stichwörter der Autor:innen

cardiac implantable electronic deviceelectrophysiology labiatrogenic complicationslead perforationpercutaneous lead removalpericardial effusiontransvenous lead extraction

Autor:innen und Institutionen

  • Cardiac Electrophysiology Section, Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio, USA.
  • Department of Thoracic and Cardiovascular Surgery, Cleveland Clinic, Cleveland, Ohio, USA.

ORCID-Profile: Ameer Awashra, Pasquale Santangeli, Thomas Callahan, Mohamad Mdaihly, Shady Nakhla, Mina K Chung

Förderung

  • The authors received no specific funding for this work.

Zitieren

APA

Younis, A., Awashra, A., Matteo, M., Hussein, A. A., Demian, J., Santangeli, P., Callahan, T., Martin, D. O., Mdaihly, M., Nakhla, S., Kanj, M., Kochar, A., Unai, S., Tong, M. Z., Sroubek, J., Saliba, W. I., Baranowski, B., Chung, M. K., Wazni, O. M., & Taigen, T. (2025). Management of Transvenous Leads in Patients With Iatrogenic Lead Perforation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70006

Vancouver

Younis A, Awashra A, Matteo M, Hussein AA, Demian J, Santangeli P, et al. Management of Transvenous Leads in Patients With Iatrogenic Lead Perforation. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.70006

BibTeX
@article{Younis2025,
  author = {Younis, Arwa and Awashra, Ameer and Matteo, Maria and Hussein, Ayman A and Demian, Joe and Santangeli, Pasquale and Callahan, Thomas and Martin, David O and Mdaihly, Mohamad and Nakhla, Shady and Kanj, Mohamed and Kochar, Arshneel and Unai, Shinya and Tong, Michael Z and Sroubek, Jakub and Saliba, Walid I and Baranowski, Bryan and Chung, Mina K and Wazni, Oussama M and Taigen, Tyler},
  title = {Management of Transvenous Leads in Patients With Iatrogenic Lead Perforation},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2025},
  doi = {10.1111/jce.70006}
}
Kostenlos starten →