Management of Transvenous Leads in Patients With Iatrogenic Lead Perforation
Journal of Cardiovascular Electrophysiology · 2025
DOI: 10.1111/jce.70006 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2025. Verfasst von Arwa Younis und 19 weiteren Autor:innen. Beteiligte Einrichtung: Cardiac Electrophysiology Section, Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio, USA. Hauptthemen laut MEDLINE: Iatrogenic Disease, Device Removal, Pacemaker, Artificial, Heart Injuries. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 3 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The authors received no specific funding for this work..
Abstract (Originalsprache)
Introduction
Iatrogenic lead perforation is a rare but serious complication of cardiac implantable electronic device (CIED) implantation. Evidence on percutaneous management of subacute or delayed cases remains limited.
Methods
We retrospectively reviewed 38 patients treated for iatrogenic lead perforation between January 2012 and October 2024. Acute lead perforation cases, which were managed during the implantation procedure, were excluded. Lead removal was performed under imaging guidance in either the electrophysiology (EP) lab or a hybrid operating room (OR) with surgical backup. Clinical presentation, imaging, procedural approach, and outcomes were analyzed.
Results
Of 38 patients, 25 underwent lead removal in the EP lab and 13 in the OR. The mean age was 67 ± 15 years; 58% were female. Most perforated leads were pacemaker leads (58%), typically involving the right ventricle (72%). Complete removal and reimplantation were achieved in all patients using simple traction with or without a locking stylet. Pericardiocentesis was required in 8% of cases. No major complications or delayed events were observed over a median 16‐month follow‐up.
Conclusion
Percutaneous removal of perforated leads is feasible and safe in the EP lab when performed by experienced operators. Surgical backup should be readily available given the potential consequences of this complication.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiac Electrophysiology Section, Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio, USA.
- Department of Thoracic and Cardiovascular Surgery, Cleveland Clinic, Cleveland, Ohio, USA.
ORCID-Profile: Ameer Awashra, Pasquale Santangeli, Thomas Callahan, Mohamad Mdaihly, Shady Nakhla, Mina K Chung
Förderung
- The authors received no specific funding for this work.
Zitieren
APA
Younis, A., Awashra, A., Matteo, M., Hussein, A. A., Demian, J., Santangeli, P., Callahan, T., Martin, D. O., Mdaihly, M., Nakhla, S., Kanj, M., Kochar, A., Unai, S., Tong, M. Z., Sroubek, J., Saliba, W. I., Baranowski, B., Chung, M. K., Wazni, O. M., & Taigen, T. (2025). Management of Transvenous Leads in Patients With Iatrogenic Lead Perforation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70006
Vancouver
Younis A, Awashra A, Matteo M, Hussein AA, Demian J, Santangeli P, et al. Management of Transvenous Leads in Patients With Iatrogenic Lead Perforation. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.70006
BibTeX
@article{Younis2025,
author = {Younis, Arwa and Awashra, Ameer and Matteo, Maria and Hussein, Ayman A and Demian, Joe and Santangeli, Pasquale and Callahan, Thomas and Martin, David O and Mdaihly, Mohamad and Nakhla, Shady and Kanj, Mohamed and Kochar, Arshneel and Unai, Shinya and Tong, Michael Z and Sroubek, Jakub and Saliba, Walid I and Baranowski, Bryan and Chung, Mina K and Wazni, Oussama M and Taigen, Tyler},
title = {Management of Transvenous Leads in Patients With Iatrogenic Lead Perforation},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2025},
doi = {10.1111/jce.70006}
}
EUVIMED