Tied up in the Lab: An Unexpected Catheter Knot in the Arterial System
Pacing and Clinical Electrophysiology · 2026-11 · 49 (11) · 1812-1815
DOI: 10.1111/pace.70287 ↗At a glance
Fallbericht in Pacing and Clinical Electrophysiology, published online on 2026-05-14. Written by Anbhigya Kumar Arya and 3 co-authors. Institution: Department of Cardiology, Vardhman Mahavir Medical College and Safdarjung Hospital, New Delhi, India. Major MEDLINE topics: Catheter Ablation, Cardiac Catheters, Ventricular Premature Complexes. 5 references.
Abstract (original language)
Catheter knotting is a rare but potentially serious complication encountered during electrophysiological procedures. We report an unusual case of catheter ablation for premature ventricular complexes (PVCs) arising from anterolateral papillary muscle, where a ThermoCool SmartTouch (ST) catheter (Biosense Webster, Irvine, CA) developed a knot within the arterial system during transaortic mapping. The knot formed after the catheter was inadvertently withdrawn into the ascending aorta and re‐advanced, leading to looping and tip entrapment. Standard maneuvers to untie the knot were unsuccessful. As a novel bailout strategy, the external portion of the catheter was cut near the handle and a large‐bore 8.5 SL0 sheath (Swartz Braided SL Transseptal Guiding Introducer Sheath; length 63 cm) was advanced over it. The knotted catheter was then maneuvered against the aortic valve, and controlled mechanical contact led to successful loosening of knot and sheath retrieval. This case underscores the importance of early recognition, careful manipulation, and innovative problem‐solving in managing rare mechanical complications in the cardiac electrophysiology lab.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Cardiology, Vardhman Mahavir Medical College and Safdarjung Hospital, New Delhi, India.
ORCID profiles: Anbhigya Kumar Arya, Anunay Gupta
Cite
APA
Arya, A. K., Gupta, A., Gupta, P., & Hs, I. (2026). Tied up in the Lab: An Unexpected Catheter Knot in the Arterial System. Pacing and Clinical Electrophysiology, 49(11), 1812-1815. https://doi.org/10.1111/pace.70287
Vancouver
Arya AK, Gupta A, Gupta P, Hs I. Tied up in the Lab: An Unexpected Catheter Knot in the Arterial System. Pacing and Clinical Electrophysiology. 2026;49(11):1812-1815. doi:10.1111/pace.70287
BibTeX
@article{Arya2026,
author = {Arya, Anbhigya Kumar and Gupta, Anunay and Gupta, Puneet and Hs, Isser},
title = {Tied up in the Lab: An Unexpected Catheter Knot in the Arterial System},
journal = {Pacing and Clinical Electrophysiology},
year = {2026},
volume = {49},
number = {11},
pages = {1812--1815},
doi = {10.1111/pace.70287}
}