Microcatheter tip fracture and acute vessel occlusion rescued with subintimal tracking and re‐entry: Move the tip
Catheterization and Cardiovascular Interventions · 2024
DOI: 10.1002/ccd.31264 ↗Kurz gefasst
Fallbericht in der Zeitschrift Catheterization and Cardiovascular Interventions, 2024. Verfasst von Francesco Giangiacomi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Sciences and Community Health, University of Milan, Milan, Italy. Hauptthemen laut MEDLINE: Cardiac Catheters, Coronary Occlusion, Angioplasty, Balloon, Coronary, Equipment Failure. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 7 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
The development of microcatheters capable of advancing through long, calcified, and tortuous vessels, produced an increase in the success rate of complex percutaneous coronary interventions. However, the aggressive manipulation of these materials may lead to potentially severe and threatening complications. We present a case of Turnpike Spiral (Teleflex) tip fracture, which caused an acute occlusion of the proximal right coronary artery. The coronary artery was tortuous and had two heavily calcified lesions in the proximal segment, uncrossable to both small angioplasty balloons and standard microcatheters. After the incarceration of the tip of the Turnpike Spiral in the calcium, it got fractured during the attempt of removing it from the vessel. This led to ST‐segment elevation and patient instability. Due to the impossibility of advancing a second guidewire into the true lumen, we decided to proceed to antegrade dissection and re‐entry using the scratch‐and‐go and subintimal tracking and re‐entry technique. We placed a guide‐extension and we used a Gaia II (Asahi Intecc) to enter the subintimal space and advance over the fractured tip. The wire was exchanged for a Gladius MG (Asahi Intecc), which re‐entered the true lumen at the level of the acute margin of the heart. Intravascular ultrasound was used to confirm the re‐entry point, and three drug eluting stents were placed. In this way we rapidly restored the flow in the right coronary artery, and we prevented a myocardial infarction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Clinical Sciences and Community Health, University of Milan, Milan, Italy.
- Department of Cardio-Thoracic-Vascular Diseases, Foundation IRCCS Ca' Granda Ospedale Maggiore Policlinico, Milan, Italy.
ORCID-Profile: Francesco Giangiacomi, Gabriele Tumminello
Zitieren
APA
Giangiacomi, F., Mircoli, L., Tumminello, G., & Carugo, S. (2024). Microcatheter tip fracture and acute vessel occlusion rescued with subintimal tracking and re‐entry: Move the tip. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31264
Vancouver
Giangiacomi F, Mircoli L, Tumminello G, Carugo S. Microcatheter tip fracture and acute vessel occlusion rescued with subintimal tracking and re‐entry: Move the tip. Catheterization and Cardiovascular Interventions. 2024. doi:10.1002/ccd.31264
BibTeX
@article{Giangiacomi2024,
author = {Giangiacomi, Francesco and Mircoli, Luca and Tumminello, Gabriele and Carugo, Stefano},
title = {Microcatheter tip fracture and acute vessel occlusion rescued with subintimal tracking and re‐entry: Move the tip},
journal = {Catheterization and Cardiovascular Interventions},
year = {2024},
doi = {10.1002/ccd.31264}
}
EUVIMED