Electrode Thermal Profile During Ventricular RF Ablation
JACC: Clinical Electrophysiology · 2026
DOI: 10.1016/j.jacep.2025.09.008 ↗Kurz gefasst
Fachartikel in der Zeitschrift JACC: Clinical Electrophysiology, 2026. Verfasst von Daisuke Togashi und 11 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Division, Department of Medicine, Vanderbilt University Medical Center, Nashville, Tennessee, USA. Hauptthemen laut MEDLINE: Catheter Ablation, Cardiac Catheters, Heart Ventricles, Tachycardia, Ventricular. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
BACKGROUND: Electrode-tissue contact and orientation influence radiofrequency (RF) ablation heating, lesion size, and steam pops, and is not easily assessed. A catheter with 6 distal electrode temperature sensors provides thermal profiles that may help clarify electrode-tissue interactions. OBJECTIVES: The aim of this study was to evaluate the relationships between thermal profiles, catheter orientation, impedance fall, steam pops, and discordant temperature-impedance changes during ventricular RF ablation. METHODS: We analyzed 2,085 RF applications in 100 patients, titrating power to a maximum temperature 10 Ω. Stable thermal profiles were classified as embedded, central, or peripheral. Catheter orientation from the force vector was classified as perpendicular, oblique, or parallel. RESULTS: The thermal profile aligned with the force vector at 60.9% of stable sites, with 72.1% of peripheral profiles being oblique/parallel and 80.4% of central profiles being perpendicular. Impedance fall was discordant with temperature at 890 (42.7%) sites. Catheter orientation and thermal profiles differed for low temperature-high impedance fall, compared with high temperature-low impedance fall discordant sites (P < 0.001). An embedded thermal profile (all sensors ≥45 °C) most often occurred with perpendicular or oblique orientations, had the greatest impedance falls, and had the highest incidence of steam pops (8.6% vs 1.3%; P < 0.001). No steam pops occurred with impedance falls
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiovascular Division, Department of Medicine, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
- Cardiovascular Division, Department of Medicine, Vanderbilt University Medical Center, Nashville, Tennessee, USA. Electronic address: william.g.stevenson@vumc.org.
Zitieren
APA
Togashi, D., Katsume, Y., Alahwany, S. H., Davogustto, G. E., Yoneda, Z. T., Richardson, T. D., Montgomery, J. A., Shen, S., Estrada, J. C., Kanagasundram, A. N., Tandri, H., & Stevenson, W. G. (2026). Electrode Thermal Profile During Ventricular RF Ablation. JACC: Clinical Electrophysiology. https://doi.org/10.1016/j.jacep.2025.09.008
Vancouver
Togashi D, Katsume Y, Alahwany SH, Davogustto GE, Yoneda ZT, Richardson TD, et al. Electrode Thermal Profile During Ventricular RF Ablation. JACC: Clinical Electrophysiology. 2026. doi:10.1016/j.jacep.2025.09.008
BibTeX
@article{Togashi2026,
author = {Togashi, Daisuke and Katsume, Yumi and Alahwany, Salah H and Davogustto, Giovanni E and Yoneda, Zachary T and Richardson, Travis D and Montgomery, Jay A and Shen, Sharon and Estrada, Juan C and Kanagasundram, Arvindh N and Tandri, Harikrishna and Stevenson, William G},
title = {Electrode Thermal Profile During Ventricular RF Ablation},
journal = {JACC: Clinical Electrophysiology},
year = {2026},
doi = {10.1016/j.jacep.2025.09.008}
}
EUVIMED