EUVIMEDEUVIMED entdecken
ÜbersichtsarbeitOpen Access

How to Use a Variable Loop Circular Catheter to Perform Zero Fluoro Pulmonary Vein Isolation With Pulsed Field Ablation

William K Chan iD, Melani Keshishi, Tara Gomez, Umjeet S Jolly

Pacing and Clinical Electrophysiology · 2026

DOI: 10.1111/pace.70244 ↗

Kurz gefasst

Übersichtsarbeit in der Zeitschrift Pacing and Clinical Electrophysiology, 2026. Verfasst von William K Chan und 3 weiteren Autor:innen. Beteiligte Einrichtung: Heart Rhythm Program, Waterloo Regional Health Network, Kitchener, Ontario, Canada. Hauptthemen laut MEDLINE: Atrial Fibrillation, Pulmonary Veins, Catheter Ablation. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The variable loop circular catheter (VLCC) called VARIPULSE is a novel device designed for catheter ablation of atrial fibrillation (AFib) and electro‐anatomical mapping via 3D intracardiac echocardiography (ICE). Described herein is a streamlined workflow for AFib ablation and reported feasibility, procedural efficiency, and early safety signals of the first 34 consecutive patients. The technical details of step‐by‐step techniques are described to serve as a practical guide for clinicians and investigators.

Methods

A retrospective, single‐center review of procedural parameters was done on 34 paroxysmal and/or persistent AFib patients who were treated with a VLCC for both mapping and ablating between January 1, 2025 and May 15, 2025. The workflow of all procedures was comprised of general anesthesia administration, placement of 2 sheaths via ultrasound guided access, administration of a heparin bolus, 3D anatomical mapping of the pulmonary veins with CARTOSOUND FAM, a transeptal puncture, VLCC ablation, and then closure. Figures illustrating key procedural steps, including catheter maneuvering, are included.

Results

All 34 cases were done without fluoroscopy, with a median procedural time of 40.5 [IQR 19] min. After 10 procedures per operator, consistently shorter procedure times (

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial FibrillationPulmonary VeinsCatheter AblationHumansRetrospective StudiesFemaleMaleIrreversible Electroporation TherapyMiddle AgedEquipment DesignFluoroscopyAgedEchocardiography, Three-DimensionalUltrasonography, Interventional

Stichwörter der Autor:innen

atrial fibrillationintracardiac echocardiogramirreversible electroporationparoxysmal atrial fibrillationpersistent atrial fibrillationpulmonary vein isolationpulsed field ablationzero fluoroscopy

Autor:innen und Institutionen

  • Heart Rhythm Program, Waterloo Regional Health Network, Kitchener, Ontario, Canada.
  • Department of Medicine, McMaster University, Hamilton, Ontario, Canada.
  • Johnson & Johnson, MedTech, 31 Technology Dr Suite, Irvine, California, USA.

ORCID-Profile: William K Chan

Zitieren

APA

Chan, W. K., Keshishi, M., Gomez, T., & Jolly, U. S. (2026). How to Use a Variable Loop Circular Catheter to Perform Zero Fluoro Pulmonary Vein Isolation With Pulsed Field Ablation. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70244

Vancouver

Chan WK, Keshishi M, Gomez T, Jolly US. How to Use a Variable Loop Circular Catheter to Perform Zero Fluoro Pulmonary Vein Isolation With Pulsed Field Ablation. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70244

BibTeX
@article{Chan2026,
  author = {Chan, William K and Keshishi, Melani and Gomez, Tara and Jolly, Umjeet S},
  title = {How to Use a Variable Loop Circular Catheter to Perform Zero Fluoro Pulmonary Vein Isolation With Pulsed Field Ablation},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2026},
  doi = {10.1111/pace.70244}
}
Kostenlos starten →