Fallbericht in der Zeitschrift Pacing and Clinical Electrophysiology, 2026. Verfasst von Kentaro Hayashi und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Arrhythmia Center, Ageo Central General Hospital, Ageo City, Saitama, Japan. Hauptthemen laut MEDLINE: Cardiac Catheters, Cryosurgery, Surgery, Computer-Assisted, Tachycardia, Supraventricular. Die Arbeit nennt 5 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Introduction
Cryoablation is widely used for septal paroxysmal supraventricular tachycardia because of its lower risk of atrioventricular block; however, real‐time quantitative assessment of catheter–tissue contact is limited. Methods and Results We describe a modified configuration enabling visualization of the contact index (CI) during cryoablation with the Freezor Xtra catheter, supported by in vitro validation and two clinical cases. In vitro testing demonstrated progressive CI elevation with decreasing catheter–tissue distance. Successful lesions were achieved at CI values of approximately 20%–25%. CI increased despite minimal electrogram changes and correlated with tissue proximity.
Conclusion
CI‐guided cryoablation is feasible and may improve contact assessment during septal ablation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Arrhythmia Center, Ageo Central General Hospital, Ageo City, Saitama, Japan.
- Department of Clinical Engineering, Ageo Central General Hospital, Ageo City, Saitama, Japan.
- Department of Cardiology, Ageo Central General Hospital, Ageo City, Saitama, Japan.
ORCID-Profile: Takeshi Kitamura
Zitieren
APA
Hayashi, K., Kitamura, T., Izumi, C., Nagahara, M., Sahashi, S., Lee, K., Sugiyama, H., Tanimoto, S., Ogata, N., & Isshiki, T. (2026). Optimization of Cryoablation Catheter–Tissue Contact Using the Contact Index. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70235
Vancouver
Hayashi K, Kitamura T, Izumi C, Nagahara M, Sahashi S, Lee K, et al. Optimization of Cryoablation Catheter–Tissue Contact Using the Contact Index. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70235
BibTeX
@article{Hayashi2026,
author = {Hayashi, Kentaro and Kitamura, Takeshi and Izumi, Chihiro and Nagahara, Masashi and Sahashi, Shuichi and Lee, Kyonghee and Sugiyama, Haruki and Tanimoto, Shuzo and Ogata, Nobuhiko and Isshiki, Takaaki},
title = {Optimization of Cryoablation Catheter–Tissue Contact Using the Contact Index},
journal = {Pacing and Clinical Electrophysiology},
year = {2026},
doi = {10.1111/pace.70235}
}
EUVIMED