Integrated dilator‐needle transseptal crossing device in atrial fibrillation cryoballoon ablation procedures
Journal of Cardiovascular Electrophysiology · 2024
DOI: 10.1111/jce.16256 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2024. Verfasst von Juan Benezet-Mazuecos und 3 weiteren Autor:innen. Beteiligte Einrichtung: Arrhythmia Unit, Department of Cardiology, Hospital Universitario La Luz, Madrid, Spain. Hauptthemen laut MEDLINE: Atrial Fibrillation, Cryosurgery, Equipment Design. Die Arbeit nennt 14 Literaturangaben.
Abstract (Originalsprache)
Introduction
In cryoballoon ablation (CBA) procedures, transseptal access (TSA) is generally achieved using a standard sheath and needle system that is exchanged for the cryoballoon delivery sheath and dilator over a long wire. Sheath exchange has been related with air embolic events. Recently, an integrated dilator‐needle system assembled to the cryoballoon sheath was introduced. We aimed to evaluate the efficacy and safety of an integrated TSA tool compared with the traditional approach in atrial fibrillation CBA procedures.
Methods
Patients scheduled for CBA procedures were randomized 1:1 to traditional TSA (t‐TSA) or integrated TSA (i‐TSA). TSA time was defined as time from superior vena cava to LA insertion of the cryoballoon delivery sheath, after sheath exchange (t‐TSA) or directly (i‐TSA).
Results
Ninety‐seven patients (76 males, mean age 59 ± 10 years) were randomized, 48 patients underwent t‐TSA, and 49 i‐TSA. Mean TSA time was 5 min 59 s ± 5 min 36 s in the t‐TSA group and 2 min 59 s ± 2 min 14 s in the i‐TSA group ( p < .001). Total fluoroscopy time, skin‐to‐skin procedure time, and LA dwell time were respectively 15 ± 6, 69 ± 16, and 44 ± 12 min in the t‐TSA group and 13 ± 6, 65 ± 15, and 43 ± 11 min in the i‐TSA group ( p = ns). No clinically significant acute complications related to TSA were noted in both cohorts.
Conclusion
This is the first randomized study comparing both TSA approaches. TSA in CBA procedures using this integrated tool enables a safe and efficient workflow, reducing TSA time and avoiding sheath exchange.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Arrhythmia Unit, Department of Cardiology, Hospital Universitario La Luz, Madrid, Spain.
ORCID-Profile: Juan Benezet-Mazuecos
Zitieren
APA
Benezet-Mazuecos, J., Lozano, Á., Miracle, Á., & Crosa, J. (2024). Integrated dilator‐needle transseptal crossing device in atrial fibrillation cryoballoon ablation procedures. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16256
Vancouver
Benezet-Mazuecos J, Lozano Á, Miracle Á, Crosa J. Integrated dilator‐needle transseptal crossing device in atrial fibrillation cryoballoon ablation procedures. Journal of Cardiovascular Electrophysiology. 2024. doi:10.1111/jce.16256
BibTeX
@article{BenezetMazuecos2024,
author = {Benezet-Mazuecos, Juan and Lozano, Álvaro and Miracle, Ángel and Crosa, Julián},
title = {Integrated dilator‐needle transseptal crossing device in atrial fibrillation cryoballoon ablation procedures},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2024},
doi = {10.1111/jce.16256}
}