Apical Lead Position Increases the Risk of Right Ventricular Lead Perforation: A Retrospective Cohort Study
Pacing and Clinical Electrophysiology · 2026
DOI: 10.1111/pace.70231 ↗Kurz gefasst
Kohortenstudie in der Zeitschrift Pacing and Clinical Electrophysiology, 2026. Verfasst von Yuji Manabe und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Yokohama Rosai Hospital, Yokohama, Japan. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Heart Ventricles, Electrodes, Implanted, Pacemaker, Artificial. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Right ventricular (RV) lead perforation is a rare but potentially life‐threatening complication of cardiac implantable electronic devices. Although apical lead placement has been proposed as a risk factor, its clinical significance remains uncertain. This study evaluated whether apical RV lead positioning increases perforation risk and characterized the clinical features, diagnosis, and outcomes of affected patients.
Methods
We retrospectively reviewed data from 1,923 patients who underwent RV lead implantation between January 2002 and June 2023. Patients were categorized into Apex and Non‐apex groups according to lead tip location. RV lead perforation cases were identified based on chest radiography findings. Baseline characteristics and the incidence and management of lead perforation were analyzed.
Results
Among the 1,923 patients, 503 (26.2%) were assigned to the Apex group and 1,420 (73.8%) to the Non‐apex group. RV lead perforation occurred in 5 patients (0.26%): 4 in the Apex group and 1 in the Non‐apex group (0.79% vs. 0.07%, p = 0.018). Perforations were diagnosed intraoperatively or within 1–3 days post‐implantation, primarily using chest computed tomography. Clinical presentations ranged from asymptomatic findings to chest pain. All patients underwent percutaneous RV lead extraction and revision, without major complications.
Conclusion
Apical RV lead positioning was significantly associated with an increased risk of lead perforation. Although perforation was uncommon, prompt diagnosis and individualized management ensured favorable outcomes. Careful attention to lead tip location during implantation may help mitigate this risk. Given the small number of events, this finding should be interpreted as hypothesis‐generating.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Yokohama Rosai Hospital, Yokohama, Japan.
ORCID-Profile: Yuji Manabe, Shinya Kowase
Zitieren
APA
Manabe, Y., Kino, T., Asano, S., Fukuzawa, T., Tanaka, S., Osada, J., Kowase, S., Aoki, H., & Yumoto, K. (2026). Apical Lead Position Increases the Risk of Right Ventricular Lead Perforation: A Retrospective Cohort Study. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70231
Vancouver
Manabe Y, Kino T, Asano S, Fukuzawa T, Tanaka S, Osada J, et al. Apical Lead Position Increases the Risk of Right Ventricular Lead Perforation: A Retrospective Cohort Study. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70231
BibTeX
@article{Manabe2026,
author = {Manabe, Yuji and Kino, Tabito and Asano, Shunichi and Fukuzawa, Tomoyuki and Tanaka, Shingo and Osada, Jun and Kowase, Shinya and Aoki, Hajime and Yumoto, Kazuhiko},
title = {Apical Lead Position Increases the Risk of Right Ventricular Lead Perforation: A Retrospective Cohort Study},
journal = {Pacing and Clinical Electrophysiology},
year = {2026},
doi = {10.1111/pace.70231}
}
EUVIMED