EUVIMEDOpdag EUVIMED
Multizentrische StudieBeobachtungsstudieOpen access

Effect of Venous Tenting Puncture Technique on Complications in Cardiac Device Implantation via Axillary Access

Emre Melik Faideci iD, Sevgi Özcan iD, Esra Dönmez iD, Murat Ziyrek iD, Mehmet Ballı iD, Ertuğrul Okuyan iD

Pacing and Clinical Electrophysiology · 2026

DOI: 10.1111/pace.70213 ↗

At a glance

Multizentrische Studie in Pacing and Clinical Electrophysiology, 2026. Written by Emre Melik Faideci and 5 co-authors. Institution: Cardiology Clinic, Istanbul Bagcilar Training and Research Hospital, Bağcılar, Istanbul, Türkiye. Major MEDLINE topics: Axillary Vein, Defibrillators, Implantable, Pacemaker, Artificial, Postoperative Complications. 28 references. The full text is freely available in PubMed Central.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Axillary vein puncture (AVP) is an effective technique in cardiac implantable electronic device (CIED) implantation. This study evaluates the venous tenting (VnT) technique as a novel method and compares its effectiveness to contrast venography (CV) in reducing early and late complications.

Methods

This retrospective multicenter study included 464 patients who underwent CIED implantation using VnT or CV techniques. Patients were matched 1:1 and divided into two groups: Group 1 ( n = 232) included those who underwent VnT, and Group 2 ( n = 232) included those who underwent CV. Early complications, including pneumothorax (PTX), pocket hematoma (PH), and pocket infection (PI), were collectively defined as total complications (TC). Late complications were defined as device lead integrity (DLI). Independent predictors of complications were identified using multivariate logistic regression analysis.

Results

The TC rate was 5.2%. VnT significantly reduced PH rates (0.9% vs. 4.3%, p = 0.019) and TC rates (2.2% vs. 8.19%, p = 0.003) compared to CV. No significant differences were observed between the groups in PTX, PI, or DLI rates. Multivariate logistic regression analysis identified VnT as an independent protective factor for both TC ( p = 0.009) and PH ( p = 0.022).

Conclusions

During CIED implantation, the VnT technique was associated with lower TC and PH than the CV technique. This study highlights the potential of VnT to improve outcomes. Further prospective studies are recommended to confirm these findings.

Subjects

MeSH terms (MEDLINE)

Axillary VeinDefibrillators, ImplantablePacemaker, ArtificialPostoperative ComplicationsProsthesis ImplantationPuncturesAgedFemaleHumansMalePhlebographyPneumothoraxRetrospective StudiesTreatment Outcome

Author keywords

axillary vein puncturecardiac electronic device implantationcontrast venographyprocedural complicationsvenous tenting technique

Authors and institutions

  • Cardiology Clinic, Istanbul Bagcilar Training and Research Hospital, Bağcılar, Istanbul, Türkiye.
  • Cardiology Clinic, Mersin City Hospital, Toroslar, Mersin, Türkiye.

ORCID profiles: Emre Melik Faideci, Sevgi Özcan, Esra Dönmez, Murat Ziyrek, Mehmet Ballı, Ertuğrul Okuyan

Cite

APA

Faideci, E. M., Özcan, S., Dönmez, E., Ziyrek, M., Ballı, M., & Okuyan, E. (2026). Effect of Venous Tenting Puncture Technique on Complications in Cardiac Device Implantation via Axillary Access. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70213

Vancouver

Faideci EM, Özcan S, Dönmez E, Ziyrek M, Ballı M, Okuyan E. Effect of Venous Tenting Puncture Technique on Complications in Cardiac Device Implantation via Axillary Access. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70213

BibTeX
@article{Faideci2026,
  author = {Faideci, Emre Melik and Özcan, Sevgi and Dönmez, Esra and Ziyrek, Murat and Ballı, Mehmet and Okuyan, Ertuğrul},
  title = {Effect of Venous Tenting Puncture Technique on Complications in Cardiac Device Implantation via Axillary Access},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2026},
  doi = {10.1111/pace.70213}
}
Start for free →