EUVIMEDEUVIMED entdecken
FallberichtOpen Access · CC BY-NC-ND 4.0

A Spinning Nightmare of Twiddler's Syndrome?

Nathan Flores Miranda, Habib Rehman Khan iD, Ahmed T Moustafa

Pacing and Clinical Electrophysiology · 2026-08 · 49 (8) · 1129-1132

DOI: 10.1111/pace.70173 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Pacing and Clinical Electrophysiology, online erschienen am 13.03.2026. Verfasst von Nathan Flores Miranda und 2 weiteren Autor:innen. Beteiligte Einrichtung: London Health Sciences Center Research Institute, Western University, London, Canada. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Equipment Failure, Electrodes, Implanted. Die Arbeit nennt 4 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Twiddler's syndrome is a known complication of CIED implantation. Most published cases of Twiddler's syndrome are described as being secondary to patient manipulation of the pulse generator leading to horizontal rotation and “reeling” of the pacemaker leads around the generator. The authors of this paper present a case of an 81‐year‐old female who had her anchoring suture across the pulse generator break after implant. This led to rotation of the pulse generator in the vertical axis and subsequently led to RV lead rotation in the longitudinal axis. The patient developed rising thresholds and low impedance on the RV lead, with fluoroscopy confirming an insulation break and lead fracture. Due to the nature of the lead rotation, extraction was impossible, and the lead was capped and buried. The case report highlights a unique mechanism by which a lead can rotate along its longitudinal axis, and is a reminder to implanting physicians regarding the importance of proper anchoring techniques.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pacemaker, ArtificialEquipment FailureElectrodes, ImplantedHumansFemaleAged, 80 and overSyndrome

Autor:innen und Institutionen

  • London Health Sciences Center Research Institute, Western University, London, Canada.

ORCID-Profile: Habib Rehman Khan

Zitieren

APA

Miranda, N. F., Khan, H. R., & Moustafa, A. T. (2026). A Spinning Nightmare of Twiddler's Syndrome?. Pacing and Clinical Electrophysiology, 49(8), 1129-1132. https://doi.org/10.1111/pace.70173

Vancouver

Miranda NF, Khan HR, Moustafa AT. A Spinning Nightmare of Twiddler's Syndrome?. Pacing and Clinical Electrophysiology. 2026;49(8):1129-1132. doi:10.1111/pace.70173

BibTeX
@article{Miranda2026,
  author = {Miranda, Nathan Flores and Khan, Habib Rehman and Moustafa, Ahmed T},
  title = {A Spinning Nightmare of Twiddler's Syndrome?},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2026},
  volume = {49},
  number = {8},
  pages = {1129--1132},
  doi = {10.1111/pace.70173}
}
Kostenlos starten →