EUVIMEDEUVIMED entdecken
Multizentrische Studie

Multicentric Experience With the Use of Short 13Fr Mechanical Rotating Dilator Sheath for Transvenous Lead Extraction

Alberto Preda, Francesco Melillo, Matteo Baroni, Alessandra Marzi, Vincenzo Schillaci, Sara Vargiu, Andrea Caccia, Fabrizio Guarracini, Lorenzo Gigli, Gabriele Paglino, Giulia Massaro, Igor Diemberger und 3 weitere

Pacing and Clinical Electrophysiology · 2025

DOI: 10.1111/pace.15146 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Pacing and Clinical Electrophysiology, 2025. Verfasst von Alberto Preda und 14 weiteren Autor:innen. Beteiligte Einrichtung: De Gasperis Cardio Center, Electrophysiology Unit, Niguarda Hospital, Milan, Italy. Hauptthemen laut MEDLINE: Device Removal, Defibrillators, Implantable, Pacemaker, Artificial, Electrodes, Implanted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 30 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The need for transvenous lead extraction (TLE) is increasing worldwide. Since the course of the subclavian vein until the junction with the superior vena cava is a frequent place of lead adherences, we aim to assess the safety and efficacy of a short 13Fr bidirectional rotational mechanical sheath as first choice. Methods and Results In this multicentric study, 202 carriers of a cardiac implantable electronic device (CIED) undergoing TLE using a short 13 Fr bidirectional rotational mechanical sheath were prospectively enrolled. All procedures were performed using a stepwise approach. The indication for TLE were infection (62%), malfunction (32%), and upgrade (6%). Overall, 471 leads were extracted: 65% pacing leads, 20% defibrillator leads, 9% coronary sinus leads, and 6% abandoned leads. Clinical success and complete lead extraction have been achieved in 97% and 95% of cases, respectively. The short sheath was always effective in gaining venous access at the start of the procedure and was sufficient for complete TLE in 67% of cases. Lead dwell time, defibrillator lead, number of leads per patient, and lead malfunction were predicting factors of long bidirectional rotational mechanical sheath use. There were no cases of intraprocedural death, and major complications were reported in 2% of patients. Overall survival was 97% at 1‐year follow‐up.

Conclusion

This multicentric experience using a short 13Fr bidirectional rotational mechanical sheath reported high safety and efficacy, demonstrating the utility of routine use of short extraction sheaths of higher caliper as the first choice.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Device RemovalDefibrillators, ImplantablePacemaker, ArtificialElectrodes, ImplantedHumansMaleFemaleAgedMiddle AgedProspective StudiesEquipment DesignEquipment FailureDevice Lead Extraction

Stichwörter der Autor:innen

TightRail Sub‐Ccomplex TLEshort bidirectional rotational mechanical sheathtranscatheter lead extraction

Autor:innen und Institutionen

  • De Gasperis Cardio Center, Electrophysiology Unit, Niguarda Hospital, Milan, Italy.
  • Clinica Montevergine, GVM Care and Research, Mercogliano, Italy.
  • Department of Cardiac Electrophysiology, IRCCS San Raffaele Scientific Institute, Milan, Italy.
  • Department of Experimental, Diagnostic and Specialty Medicine, S. Orsola Hospital, Cardio-Thoracic and Vascular Building, Bologna, Italy.
  • Cardiothoracic Department, Humanitas Gavazzeni, Bergamo, Italy.

Zitieren

APA

Preda, A., Melillo, F., Baroni, M., Marzi, A., Schillaci, V., Vargiu, S., Caccia, A., Guarracini, F., Gigli, L., Paglino, G., Massaro, G., Diemberger, I., Mascioli, G., Solimene, F., & Mazzone, P. (2025). Multicentric Experience With the Use of Short 13Fr Mechanical Rotating Dilator Sheath for Transvenous Lead Extraction. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.15146

Vancouver

Preda A, Melillo F, Baroni M, Marzi A, Schillaci V, Vargiu S, et al. Multicentric Experience With the Use of Short 13Fr Mechanical Rotating Dilator Sheath for Transvenous Lead Extraction. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.15146

BibTeX
@article{Preda2025,
  author = {Preda, Alberto and Melillo, Francesco and Baroni, Matteo and Marzi, Alessandra and Schillaci, Vincenzo and Vargiu, Sara and Caccia, Andrea and Guarracini, Fabrizio and Gigli, Lorenzo and Paglino, Gabriele and Massaro, Giulia and Diemberger, Igor and Mascioli, Giosuè and Solimene, Francesco and Mazzone, Patrizio},
  title = {Multicentric Experience With the Use of Short 13Fr Mechanical Rotating Dilator Sheath for Transvenous Lead Extraction},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2025},
  doi = {10.1111/pace.15146}
}
Kostenlos starten →