Fallbericht in Pacing and Clinical Electrophysiology, 2025. Written by Gabriele Pavani and 6 co-authors. Institution: Cardiac Electrophysiology Unit, San Luigi Gonzaga University Hospital, Orbassano, Turin, Italy. Major MEDLINE topics: Pacemaker, Artificial, Equipment Failure, Electrocardiography, Ambulatory. 11 references. The full text is freely available (open access, CC BY 4.0).

РезюмеSubjectsБиблиографски данниCiteRelated studies

Abstract (original language)

The growing prevalence of co‐implanted cardiac electronic devices increases the potential for clinically significant interactions. We describe the case of an 81‐year‐old man with a Biotronik BIOMONITOR IV implantable loop recorder (ILR) who required a Medtronic MICRA VR leadless pacemaker. Post‐implantation, interrogation with the modern Medtronic SmartSync programmer failed due to suspected electromagnetic interference from the ILR. Communication was successfully established using the older Medtronic CareLink 2090 programmer. This report highlights a unique instance of inter‐device interference, underscoring the importance of understanding programmer technology and device compatibility in complex clinical scenarios.

Subjects

MeSH terms (MEDLINE)

Pacemaker, ArtificialEquipment FailureElectrocardiography, AmbulatoryHumansMaleAged, 80 and overEquipment DesignEquipment Failure Analysis

Author keywords

Biotronik BIOMONITORMedtronic Carelink 2090Medtronic MicraMedtronic SmartSyncdevice interactionelectromagnetic compatibility (EMC)electromagnetic interference (EMI)implantable loop recorder (ILR)leadless pacemakerprogrammer communication

Authors and institutions

  • Cardiac Electrophysiology Unit, San Luigi Gonzaga University Hospital, Orbassano, Turin, Italy.
  • Interventional Cardiology Unit, San Luigi Gonzaga University Hospital, Orbassano, Turin, Italy.
  • University of Turin, Turin, Italy.

ORCID profiles: Gabriele Pavani

Cite

APA

Pavani, G., Previti, A., Garrone, P., Varalda, G., Fratini, E., Franzè, A., & Chinaglia, A. (2025). When Legacy Triumphs: Resolving Leadless Pacemaker Programming Failure Caused by Loop Recorder Interference. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70055

Vancouver

Pavani G, Previti A, Garrone P, Varalda G, Fratini E, Franzè A, et al. When Legacy Triumphs: Resolving Leadless Pacemaker Programming Failure Caused by Loop Recorder Interference. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.70055

BibTeX
@article{Pavani2025,
  author = {Pavani, Gabriele and Previti, Antonino and Garrone, Paolo and Varalda, Gianpaolo and Fratini, Emma and Franzè, Alfonso and Chinaglia, Alessandra},
  title = {When Legacy Triumphs: Resolving Leadless Pacemaker Programming Failure Caused by Loop Recorder Interference},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2025},
  doi = {10.1111/pace.70055}
}
Start for free →