When Legacy Triumphs: Resolving Leadless Pacemaker Programming Failure Caused by Loop Recorder Interference

Gabriele Pavani iD, Antonino Previti, Paolo Garrone, Gianpaolo Varalda, Emma Fratini, Alfonso Franzè, Alessandra Chinaglia

Pacing and Clinical Electrophysiology · 2025

DOI: 10.1111/pace.70055 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Pacing and Clinical Electrophysiology, 2025. Verfasst von Gabriele Pavani und 6 weiteren Autor:innen. Beteiligte Einrichtung: Cardiac Electrophysiology Unit, San Luigi Gonzaga University Hospital, Orbassano, Turin, Italy. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Equipment Failure, Electrocardiography, Ambulatory. Die Arbeit nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The growing prevalence of co‐implanted cardiac electronic devices increases the potential for clinically significant interactions. We describe the case of an 81‐year‐old man with a Biotronik BIOMONITOR IV implantable loop recorder (ILR) who required a Medtronic MICRA VR leadless pacemaker. Post‐implantation, interrogation with the modern Medtronic SmartSync programmer failed due to suspected electromagnetic interference from the ILR. Communication was successfully established using the older Medtronic CareLink 2090 programmer. This report highlights a unique instance of inter‐device interference, underscoring the importance of understanding programmer technology and device compatibility in complex clinical scenarios.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pacemaker, ArtificialEquipment FailureElectrocardiography, AmbulatoryHumansMaleAged, 80 and overEquipment DesignEquipment Failure Analysis

Stichwörter der Autor:innen

Biotronik BIOMONITORMedtronic Carelink 2090Medtronic MicraMedtronic SmartSyncdevice interactionelectromagnetic compatibility (EMC)electromagnetic interference (EMI)implantable loop recorder (ILR)leadless pacemakerprogrammer communication

Autor:innen und Institutionen

  • Cardiac Electrophysiology Unit, San Luigi Gonzaga University Hospital, Orbassano, Turin, Italy.
  • Interventional Cardiology Unit, San Luigi Gonzaga University Hospital, Orbassano, Turin, Italy.
  • University of Turin, Turin, Italy.

ORCID-Profile: Gabriele Pavani

Zitieren

APA

Pavani, G., Previti, A., Garrone, P., Varalda, G., Fratini, E., Franzè, A., & Chinaglia, A. (2025). When Legacy Triumphs: Resolving Leadless Pacemaker Programming Failure Caused by Loop Recorder Interference. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70055

Vancouver

Pavani G, Previti A, Garrone P, Varalda G, Fratini E, Franzè A, et al. When Legacy Triumphs: Resolving Leadless Pacemaker Programming Failure Caused by Loop Recorder Interference. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.70055

BibTeX
@article{Pavani2025,
  author = {Pavani, Gabriele and Previti, Antonino and Garrone, Paolo and Varalda, Gianpaolo and Fratini, Emma and Franzè, Alfonso and Chinaglia, Alessandra},
  title = {When Legacy Triumphs: Resolving Leadless Pacemaker Programming Failure Caused by Loop Recorder Interference},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2025},
  doi = {10.1111/pace.70055}
}
Kostenlos starten →