Concurrent Micra Leadless Pacemaker Implantation and AVN Ablation: Computer Modeling of Novel Risk Mitigation Strategy
Pacing and Clinical Electrophysiology · 2025
DOI: 10.1111/pace.15149 ↗At a glance
Übersichtsarbeit in Pacing and Clinical Electrophysiology, 2025. Written by Daniel Wetherbee Nelson and 6 co-authors. Institution: Aurora Cardiovascular and Thoracic Services, Aurora Sinai/Aurora St. Luke's Medical Centers, Milwaukee, Wisconsin, USA. Major MEDLINE topics: Pacemaker, Artificial, Atrial Fibrillation, Catheter Ablation, Atrioventricular Node. Cited 1 times so far (Crossref); 13 references. The full text is freely available (open access, CC BY-NC-ND 4.0).
Abstract (original language)
Background
Concurrent Micra leadless pacemaker (Medtronic, Minneapolis, Minnesota) implantation and atrioventricular node (AVN) ablation has been shown to be feasible and safe in patients with symptomatic, drug‐refractory atrial fibrillation (AF). However, major complications within the 30 days after concurrent Micra implantation and AVN ablation have been reported. We evaluated the efficacy and safety of the concurrent procedure at our institution.
Methods
We conducted a single‐center, retrospective case series of patients who underwent concurrent Micra implantation and radiofrequency (RF) AVN ablation from January 2019 to May 2023. A simulated computer model was created to characterize the interaction between the dissipated power at the Micra cathodal electrode as a function of the distance between the RF ablation catheter and the location of the return electrode.
Results
Fifteen patients were included. Most were elderly, White, female, and had persistent AF. One had transient, acute loss of ventricular capture that resulted in asystole and required emergent pacing from the ablation catheter. A proposed strategy of moving the RF return electrode to a cranial position from a caudal position was shown by computer modeling to direct more RF current away from the Micra and lower the dissipated power at the Micra cathodal electrode.
Conclusion
Concurrent Micra implantation and AVN ablation is feasible and safe and has high procedural success. An acute rise in pacing threshold can occur from RF energy, resulting in asystole. Computer modeling showed that placing the RF return electrode in the cranial position resulted in lower dissipated power at the Micra cathodal electrode.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Aurora Cardiovascular and Thoracic Services, Aurora Sinai/Aurora St. Luke's Medical Centers, Milwaukee, Wisconsin, USA.
- Academic Affairs, Cardiovascular Research, Aurora Sinai/Aurora St. Luke's Medical Centers, Milwaukee, Wisconsin, USA.
- Division of Cardiovascular Medicine, University of Wisconsin School of Medicine and Public Health, Milwaukee Clinical Campus, Milwaukee, Wisconsin, USA.
- Cardiac Rhythm Management, Medtronic, Inc., Minneapolis, Minnesota, USA.
ORCID profiles: Daniel Wetherbee Nelson, Lynn Erickson, Jodi L Zilinski, Imran K Niazi
Cite
APA
Nelson, D. W., Erickson, L., Zilinski, J. L., Zhao, Y., Karos, A., Whitman, T., & Niazi, I. K. (2025). Concurrent Micra Leadless Pacemaker Implantation and AVN Ablation: Computer Modeling of Novel Risk Mitigation Strategy. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.15149
Vancouver
Nelson DW, Erickson L, Zilinski JL, Zhao Y, Karos A, Whitman T, et al. Concurrent Micra Leadless Pacemaker Implantation and AVN Ablation: Computer Modeling of Novel Risk Mitigation Strategy. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.15149
BibTeX
@article{Nelson2025,
author = {Nelson, Daniel Wetherbee and Erickson, Lynn and Zilinski, Jodi L and Zhao, Yanzhu and Karos, Anna and Whitman, Teresa and Niazi, Imran K},
title = {Concurrent Micra Leadless Pacemaker Implantation and AVN Ablation: Computer Modeling of Novel Risk Mitigation Strategy},
journal = {Pacing and Clinical Electrophysiology},
year = {2025},
doi = {10.1111/pace.15149}
}
EUVIMED