Pulsed-field ablation for atrial fibrillation in patients with cardiac implantable electronic devices

Mohamed Abbas, Mehrdad Emami, Suraya Hani Kamsani, Jonathan P Ariyaratnam, Lauren Wilson, Simon Stolcman, Vince Schirripa, Sanjai Pattu Valappil, John Fitzgerald, Shaun Evans, Alessandra Pina, Monika Gawałko und 4 weitere

Heart Rhythm · 2025

DOI: 10.1016/j.hrthm.2025.04.045 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Heart Rhythm, 2025. Verfasst von Mohamed Abbas und 15 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Heart Rhythm Disorders, University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia. Hauptthemen laut MEDLINE: Atrial Fibrillation, Catheter Ablation, Pulmonary Veins, Defibrillators, Implantable. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 16 Literaturangaben. Gefördert von: Centre for Heart Rhythm Disorders, University of Adelaide, University of Adelaide, National Health and Medical Research Council.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Multiple studies have demonstrated the safety of pulsed-field ablation (PFA) systems; however, patients with cardiac implantable electronic devices (CIEDs) have been excluded because of concerns about the impact of strong electrical fields on device function and integrity. OBJECTIVE: This study assesses the effect of PFA on the function and integrity of CIEDs. METHODS: Patients with atrial fibrillation undergoing ablation were studied at 2 sites. PFA was performed to achieve pulmonary vein and posterior wall isolation. CIEDs were interrogated before and after PFA to assess function (threshold, sensing), integrity (impedance), and arrhythmia episodes. Real-time interrogation of the CIEDs was also performed to evaluate the effect of PFA electrical noise on the CIEDs. RESULTS: In the last 24 months, we performed 329 PFA procedures, using 4 different PFA systems; 35 of them in patients with CIEDs. The mean age was 70.0 ± 8.8 years, 40.0% female. Results of post-procedural testing revealed no significant changes in lead impedance, pacing threshold, or sensing of intrinsic activities. Results of real-time interrogation of the CIEDs revealed that PFA electrical noise was often found when the PFA catheter is in proximity (eg, right pulmonary veins and posterior wall). Moreover, PFA electrical noise resulted in ventricular pacing inhibition for 2.5 and 3.7 seconds in 2 patients and atrial pacing inhibition for 2.1 seconds in 1 patient. CONCLUSION: The function and integrity of CIEDs were not compromised by PFA in this study patient cohort; however, transient ventricular pacing inhibition was observed.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial FibrillationCatheter AblationPulmonary VeinsDefibrillators, ImplantablePacemaker, ArtificialHeart Conduction SystemHumansFemaleMaleAgedTreatment OutcomeFollow-Up StudiesMiddle Aged

Stichwörter der Autor:innen

Atrial fibrillationCardiac implantable electronic devicesCatheter ablationPulmonary vein isolationPulsed-field ablation

Autor:innen und Institutionen

  • Centre for Heart Rhythm Disorders, University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Centre for Heart Rhythm Disorders, University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia; Ashford Hospital, Adelaide, Australia.
  • Centre for Heart Rhythm Disorders, University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia; Ashford Hospital, Adelaide, Australia. Electronic address: prash.sanders@adelaide.edu.au.

Förderung

  • Centre for Heart Rhythm Disorders, University of Adelaide
  • University of Adelaide
  • National Health and Medical Research Council

Zitieren

APA

Abbas, M., Emami, M., Kamsani, S. H., Ariyaratnam, J. P., Wilson, L., Stolcman, S., Schirripa, V., Valappil, S. P., Fitzgerald, J., Evans, S., Pina, A., Gawałko, M., Jayakumar, M., Wong, C. X., Young, G. D., & Sanders, P. (2025). Pulsed-field ablation for atrial fibrillation in patients with cardiac implantable electronic devices. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2025.04.045

Vancouver

Abbas M, Emami M, Kamsani SH, Ariyaratnam JP, Wilson L, Stolcman S, et al. Pulsed-field ablation for atrial fibrillation in patients with cardiac implantable electronic devices. Heart Rhythm. 2025. doi:10.1016/j.hrthm.2025.04.045

BibTeX
@article{Abbas2025,
  author = {Abbas, Mohamed and Emami, Mehrdad and Kamsani, Suraya Hani and Ariyaratnam, Jonathan P and Wilson, Lauren and Stolcman, Simon and Schirripa, Vince and Valappil, Sanjai Pattu and Fitzgerald, John and Evans, Shaun and Pina, Alessandra and Gawałko, Monika and Jayakumar, Mohanaraj and Wong, Christopher X and Young, Glenn D and Sanders, Prashanthan},
  title = {Pulsed-field ablation for atrial fibrillation in patients with cardiac implantable electronic devices},
  journal = {Heart Rhythm},
  year = {2025},
  doi = {10.1016/j.hrthm.2025.04.045}
}
Kostenlos starten →