A Case of Very Early Lead Fracture in an Implantable Cardioverter‐Defibrillator: Management and Follow‐Up
Pacing and Clinical Electrophysiology · 2025
DOI: 10.1111/pace.15103 ↗Kurz gefasst
Fallbericht in der Zeitschrift Pacing and Clinical Electrophysiology, 2025. Verfasst von Marwan Shawki und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Austin Health, Melbourne, Australia. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Equipment Failure. Die Arbeit nennt 12 Literaturangaben.
Abstract (Originalsprache)
Introduction
Implantable cardioverter defibrillators (ICDs) lead fractures are rare but serious complications, often leading to inappropriate shocks. The early occurrence of such fractures post‐implantation is exceptionally uncommon. Methods and Results We present a case of a 53‐year‐old male with hypertrophic cardiomyopathy who experienced an inappropriate shock due to a lead fracture just 4 days after ICD implantation. The fractured lead was successfully extracted and replaced using a gooseneck snare passed through an oversized TightRail with no immediate complications.
Conclusion
Early lead fractures can occur shortly after ICD implantation, leading to inappropriate shocks. Successful management, including extraction and reimplantation, is possible even in these early cases. Clinicians should remain vigilant, and remote monitoring can play a crucial role in the timely detection of such complications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Austin Health, Melbourne, Australia.
- Faculty of Medicine, Nursing and Health Sciences, Monash University, Melbourne, Australia.
- Department of Cardiology, Royal Melbourne Hospital, Melbourne, Australia.
- Faculty of Medicine, Dentistry and Health Sciences, University of Melbourne, Melbourne, Australia.
- Department of Cardiology, Northern Health, Melbourne, Australia.
ORCID-Profile: Marwan Shawki
Zitieren
APA
Shawki, M., Mirzaee, S., McCormack, C., Wynn, G. J., & Lim, H. S. (2025). A Case of Very Early Lead Fracture in an Implantable Cardioverter‐Defibrillator: Management and Follow‐Up. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.15103
Vancouver
Shawki M, Mirzaee S, McCormack C, Wynn GJ, Lim HS. A Case of Very Early Lead Fracture in an Implantable Cardioverter‐Defibrillator: Management and Follow‐Up. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.15103
BibTeX
@article{Shawki2025,
author = {Shawki, Marwan and Mirzaee, Sam and McCormack, Cameron and Wynn, Gareth J and Lim, Han S},
title = {A Case of Very Early Lead Fracture in an Implantable Cardioverter‐Defibrillator: Management and Follow‐Up},
journal = {Pacing and Clinical Electrophysiology},
year = {2025},
doi = {10.1111/pace.15103}
}
EUVIMED