EUVIMEDEUVIMED entdecken
Crossref

Difficulty and Complications of Lead Extraction—Is ICD Lead Design Important?

Andrzej Kutarski iD, Wojciech Jacheć, Paweł Stefańczyk, Łukasz Tułecki, Dorota Nowosielecka iD

Pacing and Clinical Electrophysiology · 2026

DOI: 10.1111/pace.70144 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Pacing and Clinical Electrophysiology, 2026. Verfasst von Andrzej Kutarski und 4 weiteren Autor:innen. Beteiligte Einrichtung: 2nd Department of Cardiology, Faculty of Medical Sciences in Zabrze, Medical University of Silesia, Katowice, Poland. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Device Lead Extraction, Device Removal, Postoperative Complications. Die Arbeit nennt 35 Literaturangaben. Gefördert von: Medical University of Silesia in Katowice, Poland.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Transvenous removal of implantable cardioverter‐defibrillator (ICD) leads is considered to be more difficult and risky. The goal was to explore the impact of ICD lead design on transvenous lead extraction (TLE) safety and effectiveness.

Methods

Analysis of 3929 TLE procedures (including 1068 with ICD lead extraction) were analyzed retrospectively.

Results

The extracted ICD leads were about three years younger than pacemaker (PM) leads (medians: 61.00 vs. 97.00 months). The rate of major TLE complications (MC) was significantly higher in the PM lead extraction group (2.25% vs. 0.75%; p = 0.002) and complete procedural success was achieved significantly less often (by 4.11%; p < 0.001 in entire group of patients and by 3.74%; p < 0.001 in patients matched according to age of extracted lead). However, no significant differences were found in the scores of TLE complexity and major complications during extraction of ICD leads produced by different manufacturers. Lower rates of complete procedural success, compared to Medtronic ICD leads, were found in patients with Abbott and Biotronik ICD leads (99.45% vs. 97.32%; p = 0.009 and 96.65%; p = 0.003).

Conclusions

1. ICD lead implant duration was 3.17 years shorter and perhaps for this reason ICD lead extraction was often less difficult, the procedure less complex and MC significantly rare (0.75%) in comparison with PM lead extraction (2.55%). There were no differences between groups matched according to age of extracted lead. Achieving of complete procedural success was lower in PM lead extraction both in entire as in matched group of patients. 2. There were no significant differences in the complexity of TLE procedures or the incidence of major complications depending on ICD lead manufacturer, despite the fact that Biotronik leads were 16–19 months older. 3. The extraction of Abbot and Biotronik leads was associated with a lower rate of complete procedural success (by 2.13% and 2.80%).

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Defibrillators, ImplantableDevice Lead ExtractionDevice RemovalPostoperative ComplicationsElectrodes, ImplantedHumansFemaleMaleRetrospective StudiesMiddle AgedAgedEquipment DesignTreatment OutcomeRisk FactorsEquipment Failure

Stichwörter der Autor:innen

ICD lead extraction complexityICD lead modelscomplications of ICD lead extractiondifficulties of ICD lead extraction

Autor:innen und Institutionen

  • 2nd Department of Cardiology, Faculty of Medical Sciences in Zabrze, Medical University of Silesia, Katowice, Poland.
  • Department of Cardiology, The Pope John Paul II Province Hospital of Zamość, Zamość, Poland.
  • Department of Cardiac Surgery, The Pope John Paul II Province Hospital of Zamość, Zamość, Poland.
  • Faculty of Health Sciences, Academy of Zamość, Zamość, Poland.

ORCID-Profile: Andrzej Kutarski, Dorota Nowosielecka

Förderung

  • Medical University of Silesia in Katowice, Poland

Zitieren

APA

Kutarski, A., Jacheć, W., Stefańczyk, P., Tułecki, Ł., & Nowosielecka, D. (2026). Difficulty and Complications of Lead Extraction—Is ICD Lead Design Important?. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70144

Vancouver

Kutarski A, Jacheć W, Stefańczyk P, Tułecki Ł, Nowosielecka D. Difficulty and Complications of Lead Extraction—Is ICD Lead Design Important?. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70144

BibTeX
@article{Kutarski2026,
  author = {Kutarski, Andrzej and Jacheć, Wojciech and Stefańczyk, Paweł and Tułecki, Łukasz and Nowosielecka, Dorota},
  title = {Difficulty and Complications of Lead Extraction—Is ICD Lead Design Important?},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2026},
  doi = {10.1111/pace.70144}
}
Kostenlos starten →