EUVIMEDEUVIMED entdecken
Fallbericht

A Case of Very Early Lead Fracture in an Implantable Cardioverter‐Defibrillator: Management and Follow‐Up

Marwan Shawki iD, Sam Mirzaee, Cameron McCormack, Gareth J Wynn, Han S Lim

Pacing and Clinical Electrophysiology · 2025

DOI: 10.1111/pace.15103 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Pacing and Clinical Electrophysiology, 2025. Verfasst von Marwan Shawki und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Austin Health, Melbourne, Australia. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Equipment Failure. Die Arbeit nennt 12 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Implantable cardioverter defibrillators (ICDs) lead fractures are rare but serious complications, often leading to inappropriate shocks. The early occurrence of such fractures post‐implantation is exceptionally uncommon. Methods and Results We present a case of a 53‐year‐old male with hypertrophic cardiomyopathy who experienced an inappropriate shock due to a lead fracture just 4 days after ICD implantation. The fractured lead was successfully extracted and replaced using a gooseneck snare passed through an oversized TightRail with no immediate complications.

Conclusion

Early lead fractures can occur shortly after ICD implantation, leading to inappropriate shocks. Successful management, including extraction and reimplantation, is possible even in these early cases. Clinicians should remain vigilant, and remote monitoring can play a crucial role in the timely detection of such complications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Defibrillators, ImplantableEquipment FailureHumansMaleMiddle AgedDevice RemovalElectrodes, ImplantedCardiomyopathy, Hypertrophic

Stichwörter der Autor:innen

ICD lead fracturehypertrophic cardiomyopathyinappropriate shocklead extractionremote monitoring

Autor:innen und Institutionen

  • Department of Cardiology, Austin Health, Melbourne, Australia.
  • Faculty of Medicine, Nursing and Health Sciences, Monash University, Melbourne, Australia.
  • Department of Cardiology, Royal Melbourne Hospital, Melbourne, Australia.
  • Faculty of Medicine, Dentistry and Health Sciences, University of Melbourne, Melbourne, Australia.
  • Department of Cardiology, Northern Health, Melbourne, Australia.

ORCID-Profile: Marwan Shawki

Zitieren

APA

Shawki, M., Mirzaee, S., McCormack, C., Wynn, G. J., & Lim, H. S. (2025). A Case of Very Early Lead Fracture in an Implantable Cardioverter‐Defibrillator: Management and Follow‐Up. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.15103

Vancouver

Shawki M, Mirzaee S, McCormack C, Wynn GJ, Lim HS. A Case of Very Early Lead Fracture in an Implantable Cardioverter‐Defibrillator: Management and Follow‐Up. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.15103

BibTeX
@article{Shawki2025,
  author = {Shawki, Marwan and Mirzaee, Sam and McCormack, Cameron and Wynn, Gareth J and Lim, Han S},
  title = {A Case of Very Early Lead Fracture in an Implantable Cardioverter‐Defibrillator: Management and Follow‐Up},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2025},
  doi = {10.1111/pace.15103}
}
Kostenlos starten →