In‐vivo evaluation of catheter integrity with the use of a novel catheter torque tool
Journal of Cardiovascular Electrophysiology · 2023
DOI: 10.1111/jce.15992 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2023. Verfasst von Paige Mass und 5 weiteren Autor:innen. Beteiligte Einrichtung: Shiekh Zayed Institute for Pediatric Surgical Innovation, Children's National Hospital, Washington, District of Columbia, USA. Hauptthemen laut MEDLINE: Catheters, Catheter Ablation. Die Arbeit nennt 7 Literaturangaben. Gefördert von: FDA HHS.
Abstract (Originalsprache)
Introduction
Electrophysiology studies and ablation procedures require strength, steadiness, and dexterity to manipulate catheters. We have previously described a novel catheter torque tool (Peritorq) that improves torqueability and stability and decreases user muscle fatigue. The objective was to evaluate measures of catheter integrity with and without the torque tool in place using multiple diagnostic and ablation catheters in an adult porcine model.
Methods
Diagnostic and ablation catheters were inserted through the femoral or jugular vein into areas of the right atrium, coronary sinus (CS), and right ventricle. Electrical measurements including impedance, sensing, and capture thresholds were obtained with and without the torque tool. Ablation lesions (30 s) were given at different locations using both irrigated and nonirrigated catheters and measurements were recorded with and without the torque tool.
Results
Procedures were performed in eight adult pigs. Measurements with and without the torque tool in all locations did not differ significantly using any of the catheters. With the nonirrigated ablation catheter there was a significant difference in maximum (mean 1.7 W, p = .03) and average power (mean 9.1 W, p = .04) delivery at the PS tricuspid valve, but there were no other differences with the irrigated or nonirrigated catheters. Subjective assessment by the operator revealed a substantial improvement in maneuverability, ability to transfer torque, and stability within the cardiac space.
Conclusion
In an in‐vivo environment, a novel catheter torque tool subjectively improved catheter manipulation and did not have a significant impact on the integrity of electrophysiologic catheters. Further study including additional catheters and in‐vivo human testing is indicated.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Shiekh Zayed Institute for Pediatric Surgical Innovation, Children's National Hospital, Washington, District of Columbia, USA.
- Division of Cardiology, Children's National Hospital, Washington, District of Columbia, USA.
- Department of Mechanical Engineering, Johns Hopkins University, Baltimore, Maryland, USA.
- Biosense Webster, Diamond Bar, California, USA.
- Division of Cardiology, Montefiore Medical Center, Bronx, New York, USA.
- Division of Cardiology, Children's Hospital at Montefiore, Bronx, New York, USA.
ORCID-Profile: Luigi DiBiase, Bradley C Clark
Förderung
- FDA HHS
Zitieren
APA
Mass, P., Opfermann, J., Grupposo, R. T. V., DiBiase, L., Berul, C. I., & Clark, B. C. (2023). In‐vivo evaluation of catheter integrity with the use of a novel catheter torque tool. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.15992
Vancouver
Mass P, Opfermann J, Grupposo RTV, DiBiase L, Berul CI, Clark BC. In‐vivo evaluation of catheter integrity with the use of a novel catheter torque tool. Journal of Cardiovascular Electrophysiology. 2023. doi:10.1111/jce.15992
BibTeX
@article{Mass2023,
author = {Mass, Paige and Opfermann, Justin and Grupposo, R T Vito and DiBiase, Luigi and Berul, Charles I and Clark, Bradley C},
title = {In‐vivo evaluation of catheter integrity with the use of a novel catheter torque tool},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2023},
doi = {10.1111/jce.15992}
}
EUVIMED