Effect of contact force on pulsed field ablation lesions in porcine cardiac tissue
Journal of Cardiovascular Electrophysiology · 2023-03 · 34 (3) · 693-699
DOI: 10.1111/jce.15813 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 20.01.2023. Verfasst von Lars Mattison und 9 weiteren Autor:innen. Beteiligte Einrichtung: Medtronic, Minneapolis, Minnesota, USA. Hauptthemen laut MEDLINE: Catheter Ablation, Radiofrequency Ablation. Die Arbeit wurde bisher 83-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Introduction
Contact force has been used to titrate lesion formation for radiofrequency ablation. Pulsed field ablation (PFA) is a field‐based ablation technology for which limited evidence on the impact of contact force on lesion size is available.
Methods
Porcine hearts ( n = 6) were perfused using a modified Langendorff set‐up. A prototype focal PFA catheter attached to a force gauge was held perpendicular to the epicardium and lowered until contact was made. Contact force was recorded during each PFA delivery. Matured lesions were cross‐sectioned, stained, and the lesion dimensions measured.
Results
A total of 82 lesions were evaluated with contact forces between 1.3 and 48.6 g. Mean lesion depth was 4.8 ± 0.9 mm (standard deviation), mean lesion width was 9.1 ± 1.3 mm, and mean lesion volume was 217.0 ± 96.6 mm 3 . Linear regression curves showed an increase of only 0.01 mm in depth (depth = 0.01 × contact force + 4.41, R 2 = 0.05), 0.03 mm in width (width = 0.03 × contact force + 8.26, R 2 = 0.13) for each additional gram of contact force, and 2.20 mm 3 in volume (volume = 2.20 × contact force + 162, R 2 = 0.10).
Conclusion
Increasing contact force using a bipolar, biphasic focal PFA system has minimal effects on acute lesion dimensions in an isolated porcine heart model and achieving tissue contact is more important than the force with which that contact is made.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medtronic, Minneapolis, Minnesota, USA.
- McGill University Health Center, McGill University, Montreal, Quebec, Canada.
- Department of Cardiology, Inselspital-University Hospital Bern, University of Bern, Bern, Switzerland.
- Department of Electrophysiology, Heart Center Leipzig at University of Leipzig, Leipzig, Germany.
- Faculty of Electrical Engineering, University of Ljubljana, Ljubljana, Slovenia.
ORCID-Profile: Tobias Reichlin, Birce Önal, Daniel C Sigg
Zitieren
APA
Mattison, L., Verma, A., Tarakji, K. G., Reichlin, T., Hindricks, G., Sack, K. L., Önal, B., Schmidt, M. M., Miklavčič, D., & Sigg, D. C. (2023). Effect of contact force on pulsed field ablation lesions in porcine cardiac tissue. Journal of Cardiovascular Electrophysiology, 34(3), 693-699. https://doi.org/10.1111/jce.15813
Vancouver
Mattison L, Verma A, Tarakji KG, Reichlin T, Hindricks G, Sack KL, et al. Effect of contact force on pulsed field ablation lesions in porcine cardiac tissue. Journal of Cardiovascular Electrophysiology. 2023;34(3):693-699. doi:10.1111/jce.15813
BibTeX
@article{Mattison2023,
author = {Mattison, Lars and Verma, Atul and Tarakji, Khaldoun G and Reichlin, Tobias and Hindricks, Gerhard and Sack, Kevin L and Önal, Birce and Schmidt, Megan M and Miklavčič, Damijan and Sigg, Daniel C},
title = {Effect of contact force on pulsed field ablation lesions in porcine cardiac tissue},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2023},
volume = {34},
number = {3},
pages = {693--699},
doi = {10.1111/jce.15813}
}
EUVIMED