EUVIMEDSkopri EUVIMED
Open access · CC BY 4.0

Comprehensive Impact of Multiplanar Malalignment on Prosthetic Mechanics Under Gait Loading After Total Knee Arthroplasty—A Finite Element Analysis

Yichao Luan, Min Zhang, Xiang Dong, Hongping Duan, Zhiwei Wang, Zhichang Li iD, Cheng-Kung Cheng iD

Orthopaedic Surgery · 2025-07 · 17 (7) · 2112-2120

DOI: 10.1111/os.70068 ↗

At a glance

Journal article in Orthopaedic Surgery, published online on 2025-05-27. Written by Yichao Luan and 6 co-authors. Institution: School of Biological Science and Medical Engineering, Beihang University, Beijing, China. Major MEDLINE topics: Arthroplasty, Replacement, Knee, Knee Prosthesis, Gait, Bone Malalignment. Cited 5 times so far (Crossref); 35 references. The full text is freely available (open access, CC BY 4.0). Funded by: National Key Research and Development Program of China, Natural Science Foundation of Beijing Municipality-Haidian Original Innovation Joint Fund.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Background

Component alignment is a key factor influencing clinical outcomes after total knee arthroplasty (TKA). Previous studies have shown that single‐plane alignment can significantly affect knee joint kinematics and biomechanics. However, the comprehensive impact of multiplanar malalignment has been rarely investigated.

Objective

This study aimed to investigate the influence of the multiplanar malalignment combination on the polyethylene tibial liners under gait loading, a primary activity of daily life, as well as the degree of the influence of the alignments on the different planes.

Method

A validated finite element model of a cruciate‐retaining knee prosthesis under gait loading was used in this study. Five alignment parameters (−5°, −3°, 0°, 3°, 5°) on each plane (coronal, sagittal, and transverse) were selected to simulate clinical alignment errors, resulting in 125 models combining various alignment errors across the three planes. Boundary and loading conditions were set according to ISO 14243‐3:2014. The maximum von Mises stress and contact stress during a gait cycle were recorded for statistical analysis. A polynomial model was used for regression analysis, with the degree of influence of each alignment error on von Mises and contact stress determined by examining the quadratic coefficients.

Results

The highest Mises and contact stress values occurred with alignment errors of 5° varus, 5° flexion, and 5° internal rotation on the coronal, sagittal, and transverse planes, respectively. The lowest stress values were observed with a combination of 3° valgus, 5° flexion, and 0° internal rotation. The regression analysis yielded an R 2 value of 0.69 between alignment errors and Mises stress, with quadratic coefficients of 0.096, 0.013, and 0.064 for the coronal, sagittal, and transverse alignments, respectively. For contact stress, the R 2 was 0.697, with quadratic coefficients of 0.083, 0.002, and 0.026 for the coronal, sagittal, and transverse alignments, respectively.

Conclusion

The coronal alignment of the lower limb has the most significant impact on both Mises stress and contact stress of the tibial liner, followed by the rotational alignment of the tibial component. In contrast, the sagittal alignment of the femoral component has the least influence.

Subjects

MeSH terms (MEDLINE)

Arthroplasty, Replacement, KneeKnee ProsthesisGaitBone MalalignmentFinite Element AnalysisHumansBiomechanical PhenomenaProsthesis DesignWeight-BearingKnee Joint

Author keywords

finite element analysisgait loadingmultiplanar malalignmentprosthetic mechanicstotal knee arthroplasty

Authors and institutions

  • School of Biological Science and Medical Engineering, Beihang University, Beijing, China.
  • Beijing Naton Medical Research Institute Co., Ltd, Beijing, China.
  • Beijing Medical Implant Engineering Research Center, Beijing Naton Technology Group Co., Ltd, Beijing, China.
  • Department of Orthopaedics, Beijing Chaoyang Hospital, Capital Medical University, Beijing, China.
  • Arthritis Clinical and Research Center, Peking University People's Hospital, Beijing, China.
  • School of Biomedical Engineering, Shanghai Jiao Tong University, Engineering Research Center of Digital Medicine, Ministry of Education, Shanghai, China.

ORCID profiles: Zhichang Li, Cheng-Kung Cheng

Funding

  • National Key Research and Development Program of China
  • Natural Science Foundation of Beijing Municipality-Haidian Original Innovation Joint Fund

Cite

APA

Luan, Y., Zhang, M., Dong, X., Duan, H., Wang, Z., Li, Z., & Cheng, C. K. (2025). Comprehensive Impact of Multiplanar Malalignment on Prosthetic Mechanics Under Gait Loading After Total Knee Arthroplasty—A Finite Element Analysis. Orthopaedic Surgery, 17(7), 2112-2120. https://doi.org/10.1111/os.70068

Vancouver

Luan Y, Zhang M, Dong X, Duan H, Wang Z, Li Z, et al. Comprehensive Impact of Multiplanar Malalignment on Prosthetic Mechanics Under Gait Loading After Total Knee Arthroplasty—A Finite Element Analysis. Orthopaedic Surgery. 2025;17(7):2112-2120. doi:10.1111/os.70068

BibTeX
@article{Luan2025,
  author = {Luan, Yichao and Zhang, Min and Dong, Xiang and Duan, Hongping and Wang, Zhiwei and Li, Zhichang and Cheng, Cheng-Kung},
  title = {Comprehensive Impact of Multiplanar Malalignment on Prosthetic Mechanics Under Gait Loading After Total Knee Arthroplasty—A Finite Element Analysis},
  journal = {Orthopaedic Surgery},
  year = {2025},
  volume = {17},
  number = {7},
  pages = {2112--2120},
  doi = {10.1111/os.70068}
}
Start for free →