EUVIMEDOpdag EUVIMED
Open access · CC BY-NC 4.0

Comparison of laser speckle contrast imaging with laser Doppler perfusion imaging for tissue perfusion measurement

Goksel Guven iD, Annemieke Dijkstra, Tjallingius Martijn Kuijper, Nicole Trommel, Margriet Elisabeth van Baar, Arzu Topeli, Can Ince, Cornelis Hendrik van der Vlies

Microcirculation · 2023-01 · 30 (1)

DOI: 10.1111/micc.12795 ↗

At a glance

Journal article in Microcirculation, published online on 2022-12-29. Written by Goksel Guven and 7 co-authors. Institution: Department of Intensive Care, Erasmus MC, Rotterdam, The Netherlands. Major MEDLINE topics: Laser Speckle Contrast Imaging, Perfusion Imaging. Cited 24 times so far (Crossref); 35 references. The full text is freely available (open access, CC BY-NC 4.0). Funded by: Health~Holland, PPP Allowance, Dutch Burns Foundation.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Objective

Laser‐based tissue perfusion monitoring techniques have been increasingly used in animal and human research to assess blood flow. However, these techniques use arbitrary units, and knowledge about their comparability is scarce. This study aimed to model the relationship between laser speckle contrast imaging (LSCI) and laser Doppler perfusion imaging (LDPI), for measuring tissue perfusion over a wide range of blood flux values.

Methods

Fifteen healthy volunteers (53% female, median age 29 [IQR 22–40] years) were enrolled in this study. We performed iontophoresis with sodium nitroprusside on the forearm to induce regional vasodilation to increase skin blood flux. Besides, a stepwise vascular occlusion was applied on the contralateral upper arm to reduce blood flux. Both techniques were compared using a linear mixed model analysis.

Results

Baseline blood flux values measured by LSCI were 33 ± 6.5 arbitrary unit (AU) (Coefficient of variation [CV] = 20%) and by LDPI 60 ± 11.5 AU (CV = 19%). At the end of the iontophoresis protocol, the regional blood flux increased to 724 ± 412% and 259 ± 87% of baseline measured by LDPI and LSCI, respectively. On the other hand, during the stepwise vascular occlusion test, the blood flux reduced to 212 ± 40% and 412 ± 177% of its baseline at LDPI and LSCI, respectively. A strong correlation was found between the LSCI and LDPI instruments at increased blood flux with respect to baseline skin blood flux; however, the correlation was weak at reduced blood flux with respect to baseline.

Discussion

LSCI and LDPI instruments are highly linear for blood flux higher than baseline skin blood flux; however, the correlation decreased for blood flux lower than baseline. This study's findings could be a basis for using LSCI in specific patient populations, such as burn care.

Subjects

MeSH terms (MEDLINE)

Laser Speckle Contrast ImagingPerfusion ImagingAnimalsHumansFemaleAdultMaleBlood Flow VelocityPerfusionMicrocirculationLasersRegional Blood FlowLaser-Doppler FlowmetrySkin

Author keywords

fluxiontophoresislaser Doppler perfusion imaginglaser speckle contrast imagingmicrocirculation

Authors and institutions

  • Department of Intensive Care, Erasmus MC, Rotterdam, The Netherlands.
  • Departments of Trauma and Burn Surgery, Maasstad Ziekenhuis, Rotterdam, The Netherlands.
  • Department of Internal Medicine, Division of Intensive Care, Hacettepe University Faculty of Medicine, Ankara, Turkey.
  • Association of Dutch Burn Centers, Burn Center, Maasstad Ziekenhuis, Rotterdam, The Netherlands.
  • Trauma Research Unit Department of Surgery, Erasmus MC, Rotterdam, The Netherlands.

ORCID profiles: Goksel Guven

Funding

  • Health~Holland
  • PPP Allowance
  • Dutch Burns Foundation
  • Nederlandse Brandwonden Stichting

Cite

APA

Guven, G., Dijkstra, A., Kuijper, T. M., Trommel, N., van Baar, M. E., Topeli, A., Ince, C., & van der Vlies, C. H. (2023). Comparison of laser speckle contrast imaging with laser Doppler perfusion imaging for tissue perfusion measurement. Microcirculation, 30(1). https://doi.org/10.1111/micc.12795

Vancouver

Guven G, Dijkstra A, Kuijper TM, Trommel N, van Baar ME, Topeli A, et al. Comparison of laser speckle contrast imaging with laser Doppler perfusion imaging for tissue perfusion measurement. Microcirculation. 2023;30(1). doi:10.1111/micc.12795

BibTeX
@article{Guven2023,
  author = {Guven, Goksel and Dijkstra, Annemieke and Kuijper, Tjallingius Martijn and Trommel, Nicole and van Baar, Margriet Elisabeth and Topeli, Arzu and Ince, Can and van der Vlies, Cornelis Hendrik},
  title = {Comparison of laser speckle contrast imaging with laser Doppler perfusion imaging for tissue perfusion measurement},
  journal = {Microcirculation},
  year = {2023},
  volume = {30},
  number = {1},
  doi = {10.1111/micc.12795}
}