Changing tonometers, changing intraocular pressure? Rethinking intraocular pressure interpretation in dogs

T Yogo iD

Journal of Small Animal Practice · 2026-11-13

DOI: 10.1111/jsap.70223 ↗

At a glance

Journal article in Journal of Small Animal Practice, published online on 2026-11-13. Written by T Yogo. Institution: Laboratory of Veterinary Surgery, School of Veterinary Medicine, Nippon Veterinary and Life Science University, Tokyo, Japan. 29 references.

AbstractBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Intraocular pressure measurement is fundamental to the diagnosis and management of canine glaucoma, yet the value reported by a tonometer is inherently dependent on the instrument used. Successive generations of veterinary tonometers have improved portability, usability, standardisation and agreement with manometric pressure; however, measurements obtained with different devices are not necessarily interchangeable. Recorded intraocular pressure is also influenced by corneal properties, measurement technique, patient state and the clinical environment. These sources of variability are particularly important during longitudinal monitoring, because replacement or recalibration of a tonometer may produce an apparent change in intraocular pressure that does not reflect biological change within the eye. For practitioners, a change in tonometer should therefore be regarded as a change in the measurement system rather than simply replacement of equipment. The tonometer model and generation should be documented, and when instruments are changed, paired measurements with the old and new devices should be obtained when feasible to help establish a new clinical baseline. Population‐level inter‐device differences should not be applied as simple correction factors to individual patients, and numerical thresholds for ocular hypertension or suspicion of glaucoma should not be assumed to have equivalent clinical meaning across tonometers. When an unexpected change in intraocular pressure is detected, clinicians should consider both biological change and changes in the measurement context. Glaucoma diagnosis and longitudinal monitoring should therefore integrate intraocular pressure with visual function and structural assessment of the retina and optic nerve. Preserving the clinical meaning of intraocular pressure, rather than numerical continuity alone, should guide the interpretation of serial measurements as tonometric technology evolves.

Authors and institutions

  • Laboratory of Veterinary Surgery, School of Veterinary Medicine, Nippon Veterinary and Life Science University, Tokyo, Japan.

ORCID profiles: T Yogo

Cite

APA

Yogo, T. (2026). Changing tonometers, changing intraocular pressure? Rethinking intraocular pressure interpretation in dogs. Journal of Small Animal Practice. https://doi.org/10.1111/jsap.70223

Vancouver

Yogo T. Changing tonometers, changing intraocular pressure? Rethinking intraocular pressure interpretation in dogs. Journal of Small Animal Practice. 2026. doi:10.1111/jsap.70223

BibTeX
@article{Yogo2026,
  author = {Yogo, T},
  title = {Changing tonometers, changing intraocular pressure? Rethinking intraocular pressure interpretation in dogs},
  journal = {Journal of Small Animal Practice},
  year = {2026},
  doi = {10.1111/jsap.70223}
}
Start for free →