Reverse Torque and Wear of Internal Connections of Metal‐Based Abutments Under Cyclic Loading
Journal of Esthetic and Restorative Dentistry · 2025-06 · 37 (6) · 1539-1545
DOI: 10.1111/jerd.13425 ↗At a glance
Journal article in Journal of Esthetic and Restorative Dentistry, published online on 2025-01-29. Written by Daher Antonio Queiroz and 4 co-authors. Institution: Department of Restorative Dentistry & Prosthodontics, The University of Texas Health Science Center at Houston (UTHealth) School of Dentistry, Housto…. Major MEDLINE topics: Dental Abutments, Dental Implant-Abutment Design, Dental Restoration Wear. Cited 1 times so far (Crossref); 39 references.
Abstract (original language)
Objective
Evaluate the reverse torque values and wear of internal connections caused by metal‐based zirconia abutments under cyclic loading. Material and Methods Thirty implants were divided into three groups ( n = 10), based on connection systems: Interference‐fit cone‐screw (Cortical Master Flash MI Fit—MTA and Cortical Master Flash MI RP—MTB), and Internal hexagon (Cortical Titanium Master Connect AR—IH). Metal‐based zirconia abutments were tightened into the implant, and zirconia crowns were bonded using resin cement. All specimens underwent cyclic loading at 300 N and 9 Hz for 1,000,000 cycles. Reverse torques were assessed postloading, and wear was analyzed using scanning electron microscopy (SEM). Statistical analysis employed one‐way ANOVA for reverse torque and wear values, with significant differences assessed via the Tukey test ( α = 0.05).
Results
Significant differences in reverse torque values were found among groups: MTA (24.9 N), MTB (24.1 N), and IH (14.4 N). The IH group showed higher mean wear values on the tension side (121.0) and compression side (112.6), with a wear difference of 8.4.
Conclusion
The interference‐fit cone‐screw abutments demonstrated significantly reduced torque loss and morphology after cyclic loading, reflecting enhanced performance. Clinical Significance Although the stability of the metal base zirconia abutments to the implants was accurate, clinicians should be cautious about torque loss and more wear when using internal hexagon connections.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Department of Restorative Dentistry & Prosthodontics, The University of Texas Health Science Center at Houston (UTHealth) School of Dentistry, Houston, Texas, USA.
- Department of Prosthodontics and Periodontology, University of Campinas - Piracicaba Dental School, Piracicaba, SP, Brazil.
- Department of Prosthodontics, Tufts University School of Dental Medicine, Boston, Massachusetts, USA.
- Division of Prosthodontics, Department of Restorative Sciences, School of Dentistry, University of Minnesota, Minneapolis, Minnesota, USA.
ORCID profiles: Daher Antonio Queiroz, Mariana Itaborai Moreira Freitas, Damian Jaewhan Lee
Cite
APA
Queiroz, D. A., Freitas, M. I. M., Ockey, K. T., Lee, D. J., & Zheng, F. (2025). Reverse Torque and Wear of Internal Connections of Metal‐Based Abutments Under Cyclic Loading. Journal of Esthetic and Restorative Dentistry, 37(6), 1539-1545. https://doi.org/10.1111/jerd.13425
Vancouver
Queiroz DA, Freitas MIM, Ockey KT, Lee DJ, Zheng F. Reverse Torque and Wear of Internal Connections of Metal‐Based Abutments Under Cyclic Loading. Journal of Esthetic and Restorative Dentistry. 2025;37(6):1539-1545. doi:10.1111/jerd.13425
BibTeX
@article{Queiroz2025,
author = {Queiroz, Daher Antonio and Freitas, Mariana Itaborai Moreira and Ockey, Kaden Thomas and Lee, Damian Jaewhan and Zheng, Fengyuan},
title = {Reverse Torque and Wear of Internal Connections of Metal‐Based Abutments Under Cyclic Loading},
journal = {Journal of Esthetic and Restorative Dentistry},
year = {2025},
volume = {37},
number = {6},
pages = {1539--1545},
doi = {10.1111/jerd.13425}
}
EUVIMED