Effect of tightening protocols on abutment screw loosening and morphological changes in two-piece zirconia implants: A comparative study using cyclic torsional loading
Journal of Prosthodontic Research · 2026
DOI: 10.2186/jpr.jpr_d_25_00249 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Prosthodontic Research, 2026. Verfasst von Toma Hara und 3 weiteren Autor:innen. Beteiligte Einrichtung: Functional Occlusal Treatment, Department of Crown and Bridge Prosthodontics, The Nippon Dental University Graduate School of Life Dentistry at Niiga…. Hauptthemen laut MEDLINE: Zirconium, Dental Implants, Dental Abutments, Torsion, Mechanical. Die Arbeit nennt 49 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: This study evaluates the effects of different abutment screw tightening protocols and implant materials on screw loosening and morphological changes after cyclic torsional loading of two-piece implants. METHODS: Titanium (T) and zirconia (Z) two-piece implants were used. Abutment screws were tightened to 35 Ncm. Three protocols were tested: (A) retightening after 5 min, (B) immediate retightening, and (C) no retightening (n = 6 each). All samples underwent 100,000 cycles of cyclic torsional loading. Removal torque values before and after loading (RT-pre and RT-post) and torque loss percentage (TL) were calculated. Surface morphology was assessed using scanning electron microscopy (SEM), and dimensional changes were quantified using root mean square (RMS) values from three-dimensional data. RESULTS: In the titanium group, protocol A (TA) had lower RT-post values than protocols B (TB) and C (TC) (P < 0.001), and TB was lower than TC (P < 0.05). In the zirconia group, protocol A (ZA) was significantly lower than protocols B (ZB) and C (ZC) (P < 0.05, P < 0.001, respectively). RT-post and RT-pre were significantly lower in the zirconia group than in the titanium group under all conditions (P < 0.001). No differences were observed in TL. SEM revealed wear on the screw heads and threads in all groups. RMS values were significantly lower in the zirconia group (P < 0.05). CONCLUSIONS: Abutment screw retightening does not consistently enhance torque maintenance under torsional loading. The implant material affects torque retention and deformation, where titanium implants demonstrating greater stability than zirconia implants.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Functional Occlusal Treatment, Department of Crown and Bridge Prosthodontics, The Nippon Dental University Graduate School of Life Dentistry at Niigata, Niigata, Japan.
- Department of Crown and Bridge Prosthodontics, The Nippon Dental University School of Life Dentistry at Niigata, Niigata, Japan.
Zitieren
APA
Hara, T., Sugiki, T., Seto, M., & Ueda, K. (2026). Effect of tightening protocols on abutment screw loosening and morphological changes in two-piece zirconia implants: A comparative study using cyclic torsional loading. Journal of Prosthodontic Research. https://doi.org/10.2186/jpr.jpr_d_25_00249
Vancouver
Hara T, Sugiki T, Seto M, Ueda K. Effect of tightening protocols on abutment screw loosening and morphological changes in two-piece zirconia implants: A comparative study using cyclic torsional loading. Journal of Prosthodontic Research. 2026. doi:10.2186/jpr.jpr_d_25_00249
BibTeX
@article{Hara2026,
author = {Hara, Toma and Sugiki, Takayuki and Seto, Munetsugu and Ueda, Kazuhiko},
title = {Effect of tightening protocols on abutment screw loosening and morphological changes in two-piece zirconia implants: A comparative study using cyclic torsional loading},
journal = {Journal of Prosthodontic Research},
year = {2026},
doi = {10.2186/jpr.jpr_d_25_00249}
}