Long‐Term Performance of Single‐Chamber Atrioventricular‐Synchronous Leadless Pacing: Results From 4 Years of Serial Follow‐up

Sem Briongos-Figuero iD, Álvaro Estévez-Paniagua, Ana Sánchez-Hernández, Silvia Jiménez-Loeches, Eloy Gómez-Mariscal, Elena Sánchez-López iD, Beatriz Jiménez-Gutiérrez, Miguel Eduardo Jauregui-Abularach, David Vaqueriza-Cubillo iD, Roberto Muñoz-Aguilera

Journal of Cardiovascular Electrophysiology · 2026

DOI: 10.1111/jce.70490 ↗

At a glance

Journal article in Journal of Cardiovascular Electrophysiology, 2026. Written by Sem Briongos-Figuero and 9 co-authors. Institution: Cardiology Department, Hospital Infanta Leonor, Madrid, Spain. Major MEDLINE topics: Cardiac Pacing, Artificial, Heart Rate, Atrioventricular Node, Arrhythmias, Cardiac. 15 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Introduction

Micra AV enables atrioventricular (AV)–synchronous pacing by sensing atrial mechanical contraction (A4) using device's accelerometer. Short‐term studies showed good AV synchrony (AVS), but long‐term data are lacking.

Methods

Within the OPTIVALL cohort, we identified patients implanted with a first‐generation Micra AV (June 2020–February 2024) who remained in sinus rhythm and programmed in VDD for ≥ 1 year. Device‐derived AVS was collected annually (tracking index and total AVS = AM‐VP + AM‐VS) up to 4 years. A4 amplitude was recorded at each visit. A subgroup underwent 24‐h Holter‐ECG to assess ambulatory AVS at long‐term follow‐up. Atrial sensing had been optimized early after implant with no further adjustment at follow‐up.

Results

Of 54 implanted patients, 44 met inclusion criteria. Median (IQR) total AVS remained stable over time: 90.6% (85.4%–93.6%), 90.6% (83.1%–92.1%), 89.2% (84.3%–92.2%), and 89.3% (86.2%–91.6%) at 1, 2, 3, and 4 years, respectively ( p = 0.789). Median (IQR) tracking index was 94.3% (88.9%–97.1%), 93.0% (85.0%–98.0%), 90.2% (84.4%–98.0%), and 90.0% (87.0%–93.9%) ( p = 0.059). Median (IQR) A4 signal amplitude was 3.1 (2.0–5.1), 5.1 (2.1–5.1), 3.7 (1.9–5.1), and 5.1 (2.5–5.1) m/s 2 ( p = 0.993). In 20 patients with 24‐h Holter‐ECG at 2.7 ± 1.1 years, ambulatory AVS was 85.5% (80.0%–90.7%) and was comparable to the 3‐month value (87.9% (76.8%–93.1%); p = 0.811).

Conclusions

When atrial sensing is optimized early, Micra AV can provide high and stable AV‐synchronous pacing for up to 4 years, with preserved A4 signal amplitude.

Subjects

MeSH terms (MEDLINE)

Cardiac Pacing, ArtificialHeart RateAtrioventricular NodeArrhythmias, CardiacHumansTime FactorsFemaleMaleTreatment OutcomeAgedElectrocardiography, AmbulatoryAction PotentialsMiddle AgedFollow-Up Studies

Author keywords

A4 signalatrioventricular synchronylong‐term follow‐upprogrammingsingle chamber leadless pacing

Authors and institutions

  • Cardiology Department, Hospital Infanta Leonor, Madrid, Spain.
  • Cardiology Department, Hospital 12 de Octubre, Madrid, Spain.

ORCID profiles: Sem Briongos-Figuero, Elena Sánchez-López, David Vaqueriza-Cubillo

Cite

APA

Briongos-Figuero, S., Estévez-Paniagua, Á., Sánchez-Hernández, A., Jiménez-Loeches, S., Gómez-Mariscal, E., Sánchez-López, E., Jiménez-Gutiérrez, B., Jauregui-Abularach, M. E., Vaqueriza-Cubillo, D., & Muñoz-Aguilera, R. (2026). Long‐Term Performance of Single‐Chamber Atrioventricular‐Synchronous Leadless Pacing: Results From 4 Years of Serial Follow‐up. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70490

Vancouver

Briongos-Figuero S, Estévez-Paniagua Á, Sánchez-Hernández A, Jiménez-Loeches S, Gómez-Mariscal E, Sánchez-López E, et al. Long‐Term Performance of Single‐Chamber Atrioventricular‐Synchronous Leadless Pacing: Results From 4 Years of Serial Follow‐up. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70490

BibTeX
@article{BriongosFiguero2026,
  author = {Briongos-Figuero, Sem and Estévez-Paniagua, Álvaro and Sánchez-Hernández, Ana and Jiménez-Loeches, Silvia and Gómez-Mariscal, Eloy and Sánchez-López, Elena and Jiménez-Gutiérrez, Beatriz and Jauregui-Abularach, Miguel Eduardo and Vaqueriza-Cubillo, David and Muñoz-Aguilera, Roberto},
  title = {Long‐Term Performance of Single‐Chamber Atrioventricular‐Synchronous Leadless Pacing: Results From 4 Years of Serial Follow‐up},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2026},
  doi = {10.1111/jce.70490}
}
Start for free →