EUVIMEDDécouvrir EUVIMED
Open access · CC BY 4.0

In Vitro Assessment of 3T MRI Effects on Non–3T‐Compatible Cardiac Implantable Electronic Devices

Yu Fukuoka iD, Yasuo Sakurai, Risa Kanai, Yuuya Miyazaki, Haruka Yamazaki, Yusuke Koga, Isao Tsukamoto, Yoshifumi Ikeda, Hitoshi Mori iD, Kazuhisa Matsumoto, Ritsushi Kato iD

Journal of Cardiovascular Electrophysiology · 2026-08 · 37 (8) · 1664-1674

DOI: 10.1111/jce.70397 ↗

At a glance

Journal article in Journal of Cardiovascular Electrophysiology, published online on 2026-06-07. Written by Yu Fukuoka and 10 co-authors. Institution: Clinical Engineering Department, Saitama Medical University International Medical Center, Hidaka, Saitama, Japan. Major MEDLINE topics: Pacemaker, Artificial, Magnetic Resonance Imaging, Defibrillators, Implantable. 17 references. The full text is freely available (open access, CC BY 4.0).

RésuméSubjectsRéférences bibliographiquesCiteRelated studies

Abstract (original language)

Background

The safety of cardiac implantable electronic devices (CIEDs) during 3‐Tesla (3T) MRI remains uncertain, particularly for devices not designed for 3T use. We experimentally evaluated 3T MRI effects on 3T‐compatible and non–3T‐compatible CIEDs.

Methods

Fifty‐eight CIEDs (37 pacemakers, 21 ICDs; 3T MRI‐conditional n = 31, 1.5‐T MRI‐conditional n = 14, and non–MRI‐conditional n = 13) were tested in a 3T scanner. Each device was immersed in saline with connected leads and programmed to asynchronous or demand pacing with MRI mode ON or OFF. Device interrogation before and after scanning assessed telemetry integrity, parameter retention, and pacing‐mode changes.

Results

Most CIEDs completed MRI without malfunction and remained telemetric. However, among pacemakers, telemetry failure or parameter reset occurred in 6 non–MRI‐conditional and 3 1.5‐T MRI‐conditional devices, including some devices with MRI mode ON. Reset or telemetry failure was more frequent in devices with low battery status (≤ 30% to ERI) than in those with higher status (6/11 [55%] vs. 3/22 [14%], p = 0.013). With MRI mode OFF, magnet mode was activated in 11 3T and 7 non–MRI‐conditional pacemakers; activation frequency did not differ between asynchronous and demand pacing (16/31 [52%] vs. 18/31 [58%], p = 0.610). In some pacemakers and ICDs, demand pacing led to atrial mode switch or pacing inhibition, but no reset or telemetry failure occurred in ICDs.

Conclusions

In direct in vitro 3T MRI exposure experiments involving non–3T MRI CIEDs, some pacemakers approaching ERI demonstrated an increased risk of telemetry failure or reset, and those with MRI mode turned off showed non‐negligible mode switching. However, the majority of devices did not exhibit severe electrical malfunctions necessitating replacement in a clinical setting, and no instances of device reset or telemetry failure were observed in ICDs.

Subjects

MeSH terms (MEDLINE)

Pacemaker, ArtificialMagnetic Resonance ImagingDefibrillators, ImplantableTelemetryHumansEquipment DesignEquipment Failure Analysis

Author keywords

3 Tesla MRIcardiac implantable electronic devicemagnet modepacemakertelemetry failure

Authors and institutions

  • Clinical Engineering Department, Saitama Medical University International Medical Center, Hidaka, Saitama, Japan.
  • Central Radiology Department, Saitama Medical University, International Medical Center, Hidaka, Saitama, Japan.
  • Outpatient Clinic of Cardiology, Saitama Medical University, International Medical Center, Hidaka, Saitama, Japan.
  • Department of Cardiology, Saitama Medical University, International Medical Center, Hidaka, Saitama, Japan.

ORCID profiles: Yu Fukuoka, Hitoshi Mori, Ritsushi Kato

Cite

APA

Fukuoka, Y., Sakurai, Y., Kanai, R., Miyazaki, Y., Yamazaki, H., Koga, Y., Tsukamoto, I., Ikeda, Y., Mori, H., Matsumoto, K., & Kato, R. (2026). In Vitro Assessment of 3T MRI Effects on Non–3T‐Compatible Cardiac Implantable Electronic Devices. Journal of Cardiovascular Electrophysiology, 37(8), 1664-1674. https://doi.org/10.1111/jce.70397

Vancouver

Fukuoka Y, Sakurai Y, Kanai R, Miyazaki Y, Yamazaki H, Koga Y, et al. In Vitro Assessment of 3T MRI Effects on Non–3T‐Compatible Cardiac Implantable Electronic Devices. Journal of Cardiovascular Electrophysiology. 2026;37(8):1664-1674. doi:10.1111/jce.70397

BibTeX
@article{Fukuoka2026,
  author = {Fukuoka, Yu and Sakurai, Yasuo and Kanai, Risa and Miyazaki, Yuuya and Yamazaki, Haruka and Koga, Yusuke and Tsukamoto, Isao and Ikeda, Yoshifumi and Mori, Hitoshi and Matsumoto, Kazuhisa and Kato, Ritsushi},
  title = {In Vitro Assessment of 3T MRI Effects on Non–3T‐Compatible Cardiac Implantable Electronic Devices},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2026},
  volume = {37},
  number = {8},
  pages = {1664--1674},
  doi = {10.1111/jce.70397}
}
Start for free →