An Interchangeability Test Method for Developing IPG Impedance Equivalency Evidence for Cardiac Implantable Electronic Device Under MRI
Magnetic Resonance in Medicine · 2026-08 · 96 (2) · 883-891
DOI: 10.1002/mrm.70389 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 26.04.2026. Verfasst von Hongbae Jeong und 2 weiteren Autor:innen. Beteiligte Einrichtung: Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, Marylan…. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Pacemaker, Artificial, Defibrillators, Implantable. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Patients with cardiac implantable electronic devices (CIEDs) may receive replacement device components over time, resulting in mixed‐manufacturer systems that pose unknown, potentially higher risk under MRI exposure. It is not clear how to efficiently evaluate whether adding a non‐original implantable pulse generator (IPG) changes the system's radiofrequency (RF)–induced heating response. This study aimed to develop a least‐burdensome RF safety evaluation method for MR Conditional labeling of CIEDs. Critical parameters were characterized and demonstrated as an IPG equivalency test method that supports leveraging existing single‐manufacturer MR‐Conditional labeling for mixed‐manufacturer systems.
Methods
The transfer function model of the pacing lead was measured using the original system (Lead A + IPG A ) or a third‐party IPGs (Lead A + IPG B or IPG C ). Deposited power near the lead tip was compared for 50 Ω and open‐end conditions, and lead or IPG impedances were measured at 64 and 128 MHz.
Results
Lead tip heating was not significantly different between original and third‐party IPGs configurations ( p < 0.001). RF‐induced heating responses were outside of measurement uncertainty when the lead was connected with a 50 Ω resistor and some open‐end conditions. The results showed that small changes in IPG impedance (< 2.5 Ω) to the lead‐IPG system impedance (> 80.2 Ω) may not significantly alter RF‐induced heating, remaining within measurement uncertainties.
Conclusion
A simplified IPG impedance equivalency test has been developed to facilitate MRI access for patients with CIEDs that undergo IPG interchange. This test aids assessment of whether replacing an original IPG model with a different model may significantly increase RF‐induced heating compared to the original system.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, Maryland, USA.
- Office of Cardiovascular Devices, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, Maryland, USA.
ORCID-Profile: Hongbae Jeong
Zitieren
APA
Jeong, H., Smith, A., & Kumar, A. (2026). An Interchangeability Test Method for Developing IPG Impedance Equivalency Evidence for Cardiac Implantable Electronic Device Under MRI. Magnetic Resonance in Medicine, 96(2), 883-891. https://doi.org/10.1002/mrm.70389
Vancouver
Jeong H, Smith A, Kumar A. An Interchangeability Test Method for Developing IPG Impedance Equivalency Evidence for Cardiac Implantable Electronic Device Under MRI. Magnetic Resonance in Medicine. 2026;96(2):883-891. doi:10.1002/mrm.70389
BibTeX
@article{Jeong2026,
author = {Jeong, Hongbae and Smith, Andrew and Kumar, Ananda},
title = {An Interchangeability Test Method for Developing IPG Impedance Equivalency Evidence for Cardiac Implantable Electronic Device Under MRI},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {2},
pages = {883--891},
doi = {10.1002/mrm.70389}
}
EUVIMED