In Vitro Assessment of 3T MRI Effects on Non–3T‐Compatible Cardiac Implantable Electronic Devices
Journal of Cardiovascular Electrophysiology · 2026-08 · 37 (8) · 1664-1674
DOI: 10.1111/jce.70397 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 07.06.2026. Verfasst von Yu Fukuoka und 10 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Engineering Department, Saitama Medical University International Medical Center, Hidaka, Saitama, Japan. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Magnetic Resonance Imaging, Defibrillators, Implantable. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The safety of cardiac implantable electronic devices (CIEDs) during 3‐Tesla (3T) MRI remains uncertain, particularly for devices not designed for 3T use. We experimentally evaluated 3T MRI effects on 3T‐compatible and non–3T‐compatible CIEDs.
Methods
Fifty‐eight CIEDs (37 pacemakers, 21 ICDs; 3T MRI‐conditional n = 31, 1.5‐T MRI‐conditional n = 14, and non–MRI‐conditional n = 13) were tested in a 3T scanner. Each device was immersed in saline with connected leads and programmed to asynchronous or demand pacing with MRI mode ON or OFF. Device interrogation before and after scanning assessed telemetry integrity, parameter retention, and pacing‐mode changes.
Results
Most CIEDs completed MRI without malfunction and remained telemetric. However, among pacemakers, telemetry failure or parameter reset occurred in 6 non–MRI‐conditional and 3 1.5‐T MRI‐conditional devices, including some devices with MRI mode ON. Reset or telemetry failure was more frequent in devices with low battery status (≤ 30% to ERI) than in those with higher status (6/11 [55%] vs. 3/22 [14%], p = 0.013). With MRI mode OFF, magnet mode was activated in 11 3T and 7 non–MRI‐conditional pacemakers; activation frequency did not differ between asynchronous and demand pacing (16/31 [52%] vs. 18/31 [58%], p = 0.610). In some pacemakers and ICDs, demand pacing led to atrial mode switch or pacing inhibition, but no reset or telemetry failure occurred in ICDs.
Conclusions
In direct in vitro 3T MRI exposure experiments involving non–3T MRI CIEDs, some pacemakers approaching ERI demonstrated an increased risk of telemetry failure or reset, and those with MRI mode turned off showed non‐negligible mode switching. However, the majority of devices did not exhibit severe electrical malfunctions necessitating replacement in a clinical setting, and no instances of device reset or telemetry failure were observed in ICDs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Clinical Engineering Department, Saitama Medical University International Medical Center, Hidaka, Saitama, Japan.
- Central Radiology Department, Saitama Medical University, International Medical Center, Hidaka, Saitama, Japan.
- Outpatient Clinic of Cardiology, Saitama Medical University, International Medical Center, Hidaka, Saitama, Japan.
- Department of Cardiology, Saitama Medical University, International Medical Center, Hidaka, Saitama, Japan.
ORCID-Profile: Yu Fukuoka, Hitoshi Mori, Ritsushi Kato
Zitieren
APA
Fukuoka, Y., Sakurai, Y., Kanai, R., Miyazaki, Y., Yamazaki, H., Koga, Y., Tsukamoto, I., Ikeda, Y., Mori, H., Matsumoto, K., & Kato, R. (2026). In Vitro Assessment of 3T MRI Effects on Non–3T‐Compatible Cardiac Implantable Electronic Devices. Journal of Cardiovascular Electrophysiology, 37(8), 1664-1674. https://doi.org/10.1111/jce.70397
Vancouver
Fukuoka Y, Sakurai Y, Kanai R, Miyazaki Y, Yamazaki H, Koga Y, et al. In Vitro Assessment of 3T MRI Effects on Non–3T‐Compatible Cardiac Implantable Electronic Devices. Journal of Cardiovascular Electrophysiology. 2026;37(8):1664-1674. doi:10.1111/jce.70397
BibTeX
@article{Fukuoka2026,
author = {Fukuoka, Yu and Sakurai, Yasuo and Kanai, Risa and Miyazaki, Yuuya and Yamazaki, Haruka and Koga, Yusuke and Tsukamoto, Isao and Ikeda, Yoshifumi and Mori, Hitoshi and Matsumoto, Kazuhisa and Kato, Ritsushi},
title = {In Vitro Assessment of 3T MRI Effects on Non–3T‐Compatible Cardiac Implantable Electronic Devices},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2026},
volume = {37},
number = {8},
pages = {1664--1674},
doi = {10.1111/jce.70397}
}
EUVIMED