Pulsed Field Versus Radiofrequency Ablation in Concomitant Pulmonary Vein Isolation and LAA Occlusion: Impact on Peridevice Leak and Procedural Efficiency
Journal of Cardiovascular Electrophysiology · 2026
DOI: 10.1111/jce.70286 ↗At a glance
Vergleichsstudie in Journal of Cardiovascular Electrophysiology, 2026. Written by Mohammad Montaser Atasi and 23 co-authors. Institution: Tulane Research Innovation for Arrhythmia Discovery, Tulane University, New Orleans, Louisiana, USA. Major MEDLINE topics: Atrial Fibrillation, Pulmonary Veins, Left Atrial Appendage Closure, Catheter Ablation. Cited 4 times so far (Crossref); 29 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Background
Concomitant pulmonary vein isolation (PVI) and left atrial appendage occlusion (LAAO) is increasingly performed in patients with atrial fibrillation (AF) at elevated thromboembolic risk. Radiofrequency ablation (RFA) has been the conventional approach, but pulsed field ablation (PFA) may reduce tissue edema, procedural duration, and peridevice leak (PDL).
Objective
To compare safety, efficacy, and PDL rates between PFA + LAAO and RFA + LAAO.
Methods
In this single‐center, retrospective study, 175 consecutive patients undergoing LAAO with or without concomitant PVI between September 2021 and May 2025 were included: RFA + LAAO ( n = 61), PFA + LAAO ( n = 39), and LAAO‐only ( n = 75). Procedural data, device characteristics, and complications were collected. The primary endpoint was PDL at 45‐day follow‐up transesophageal echocardiography (TEE). Secondary endpoints included procedure duration, vascular and esophageal complications, device‐related thrombosis, and arrhythmia recurrence.
Results
Baseline characteristics were generally balanced, although RFA + LAAO patients were younger. No large PDL (> 5 mm) was observed. At 45 days, small PDL (≤ 5 mm) was significantly higher in RFA + LAAO (23%) vs. PFA + LAAO (5.1%) and LAAO‐only (9.3%; p = 0.015). Mean procedure duration was shorter with PFA + LAAO (67.1 ± 10.4 min) compared with RFA + LAAO (99.6 ± 12.8 min; p < 0.001). Major complications and device‐related thrombosis were rare. Minor complications and AF recurrence were comparable across groups.
Conclusions
Concomitant PFA + LAAO is associated with lower PDL incidence and shorter procedural times compared with RFA + LAAO, without compromising safety.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Tulane Research Innovation for Arrhythmia Discovery, Tulane University, New Orleans, Louisiana, USA.
ORCID profiles: Mohammad Montaser Atasi, Christian M Massad, Ghassan Bidaoui, Hadi Younes, Omar Kreidieh, Nassir F Marrouche
Cite
APA
Atasi, M. M., Hubbard, S., Campbell, C., Noujaim, C., Massad, C. M., Khalil, M. A., Menassa, Y., El Darzi, A., Moersdorf, M., El Khoury, C., Lim, C., Liu, Y., Jia, Y., Hassan, A., Bsoul, M., Bidaoui, G., Younes, H., Young, M., Wanna, B., …, & Marrouche, N. F. (2026). Pulsed Field Versus Radiofrequency Ablation in Concomitant Pulmonary Vein Isolation and LAA Occlusion: Impact on Peridevice Leak and Procedural Efficiency. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70286
Vancouver
Atasi MM, Hubbard S, Campbell C, Noujaim C, Massad CM, Khalil MA, et al. Pulsed Field Versus Radiofrequency Ablation in Concomitant Pulmonary Vein Isolation and LAA Occlusion: Impact on Peridevice Leak and Procedural Efficiency. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70286
BibTeX
@article{Atasi2026,
author = {Atasi, Mohammad Montaser and Hubbard, Shea and Campbell, Charles and Noujaim, Charbel and Massad, Christian M and Khalil, Michel Abou and Menassa, Yara and El Darzi, Alex and Moersdorf, Maximilian and El Khoury, Carlo and Lim, Chanho and Liu, Yingshuo and Jia, Yishi and Hassan, Abboud and Bsoul, Mayana and Bidaoui, Ghassan and Younes, Hadi and Young, Monique and Wanna, Bassam and Pandey, Amitabh C and Kreidieh, Omar and Feng, Han and Marashly, Qussay and Marrouche, Nassir F},
title = {Pulsed Field Versus Radiofrequency Ablation in Concomitant Pulmonary Vein Isolation and LAA Occlusion: Impact on Peridevice Leak and Procedural Efficiency},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2026},
doi = {10.1111/jce.70286}
}
EUVIMED