Validierungsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2026. Verfasst von Ryosuke Kato und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Medicine, Institute of Science Tokyo Hospital, Tokyo, Japan. Hauptthemen laut MEDLINE: Cardiac Catheters, Magnetics, Catheter Ablation. Die Arbeit nennt 10 Literaturangaben.
Abstract (Originalsprache)
Introduction
Most multipolar mapping catheters use a proximal magnetic sensor, risking distal positional error and slower point acquisition. We tested whether adding distal sensors improves speed and geometric fidelity vs. a conventional design.
Methods
In a preclinical swine model, maps were acquired on EnSite X. For each region—RA, RV, LA, LV, coronary sinus (CS), and epicardium (Epi)—two maps were created in sequence (HD Grid X first, then HD Grid) and mapping time was recorded. To evaluate geometric accuracy, we sampled 12 predefined sites per chamber, arranged as 3 longitudinal levels (proximal, mid, distal) × 4 anatomical lines (anterior, lateral, posterior, septal). At each site, the orthogonal catheter‐to‐surface distance was measured on the HD Grid X map using a TACTIFLEX Ablation Catheter.
Results
Across 33 paired maps (9 swine), mapping time was shorter with HD Grid X: 11.5 [8.1–13.6] vs. 15.3 [10.9–17.3] min; p < 0.001 (−24.8%), with significant reductions in RA, RV, and LV. In 243 paired points (5 swine), geometric accuracy was superior with HD Grid X: 0.5 [0.2–1.3] vs. 2.9 [2.1–4.2] mm; p < 0.001, consistent across chambers and sub‐locations.
Conclusion
Distal magnetic sensors enable faster and more faithful geometry vs. a proximal‐sensor design.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Cardiovascular Medicine, Institute of Science Tokyo Hospital, Tokyo, Japan.
- Department of Cardiovascular Medicine, Akita University Graduate School of Medicine, Akita, Japan.
- Division of Advanced Arrhythmia Research, Institute of Science Tokyo, Tokyo, Japan.
- Japan Small Animal Medical Center, Saitama, Japan.
- Animal Research Facilities, Bioscience Center, Research Infrastructure Management Center, Institute of Science Tokyo, Tokyo, Japan.
ORCID-Profile: Ryosuke Kato, Masateru Takigawa, Hidehiro Iwakawa, Shinsuke Miyazaki, Hiroyuki Watanabe
Zitieren
APA
Kato, R., Takigawa, M., Kawamura, I., Honda, M., Negishi, M., Tateishi, R., Iwakawa, H., Yamao, K., Goto, K., Nishimura, T., Tao, S., Sayaka Suzuki, D. V. M., Iwanaga, T., Miyazaki, S., Watanabe, H., & Sasano, T. (2026). Distal Magnetic Sensors Eliminate Geometric Inaccuracy and Accelerate Mapping: A Preclinical Validation of a Novel Grid‐Style Catheter. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70284
Vancouver
Kato R, Takigawa M, Kawamura I, Honda M, Negishi M, Tateishi R, et al. Distal Magnetic Sensors Eliminate Geometric Inaccuracy and Accelerate Mapping: A Preclinical Validation of a Novel Grid‐Style Catheter. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70284
BibTeX
@article{Kato2026,
author = {Kato, Ryosuke and Takigawa, Masateru and Kawamura, Iwanari and Honda, Masaki and Negishi, Miho and Tateishi, Ryo and Iwakawa, Hidehiro and Yamao, Kazuya and Goto, Kentaro and Nishimura, Takuro and Tao, Susumu and Sayaka Suzuki, D V M and Iwanaga, Takehiro and Miyazaki, Shinsuke and Watanabe, Hiroyuki and Sasano, Tetsuo},
title = {Distal Magnetic Sensors Eliminate Geometric Inaccuracy and Accelerate Mapping: A Preclinical Validation of a Novel Grid‐Style Catheter},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2026},
doi = {10.1111/jce.70284}
}
EUVIMED