Feasibility of a Cryo‐Anchoring Catheter for Linear Ablation
Journal of Cardiovascular Electrophysiology · 2025
DOI: 10.1111/jce.70131 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2025. Verfasst von Jaspal Singh Gill und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, St. George's University Hospitals NHS Foundation Trust, London, UK. Hauptthemen laut MEDLINE: Cryosurgery, Cardiac Catheters, Catheter Ablation. Die Arbeit nennt 19 Literaturangaben. Gefördert von: The authors received no specific funding for this work..
Abstract (Originalsprache)
Background
Linear ablation lesions are used in multiple ablation strategies including left atrial roof lines, and mitral and tricuspid isthmus lines. A continuous and transmural line is mandatory for effective lesions. A proposed catheter employs cryo‐anchoring: the tip initially freezes to adhere to the tissue, permitting the shaft of the catheter to be applied firmly against tissue to create linear lesions.
Methods
Ex vivo tests were performed in a 35°C water bath on turkey skeletal muscle, lamb, and porcine myocardial tissue. The experimental setup allowed the quantification of force exerted from the catheter onto the tissue with bidirectional strain‐gauge measurement. Five protocols evaluated: force for tissue or equipment failure during traction, consistency experiments to evaluate the safe level of traction, force for tissue failure in forward pressure, contact force effect on time to transmural lesion, and contact force effect on lesion depth.
Results
Tissue failure in traction occurred at the lowest force of 3.9 kg. This occurred at the ventricular apex of a lamb heart, whereas atrial samples tolerated > 10 kg of force before failure. Correlations between force, lesion thickness, and time to the formation of a transmural lesion were observed. Lesions of 3−4 mm depth occurred at 45 g forward pressure with 60 s lesions. The time required to attain this depth fell to 20 s with forces over 200 g.
Conclusion
There is a broad safety margin between the force required to enhance lesion formation during cryotherapy and the force that can disrupt tissue, both for traction and forward pressure.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, St. George's University Hospitals NHS Foundation Trust, London, UK.
ORCID-Profile: Jaspal Singh Gill, Viral Sagar, Lisa Leung, Mark M Gallagher
Förderung
- The authors received no specific funding for this work.
Zitieren
APA
Gill, J. S., Sagar, V., Leung, L., Akhtar, Z., Zuberi, Z., & Gallagher, M. M. (2025). Feasibility of a Cryo‐Anchoring Catheter for Linear Ablation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70131
Vancouver
Gill JS, Sagar V, Leung L, Akhtar Z, Zuberi Z, Gallagher MM. Feasibility of a Cryo‐Anchoring Catheter for Linear Ablation. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.70131
BibTeX
@article{Gill2025,
author = {Gill, Jaspal Singh and Sagar, Viral and Leung, Lisa and Akhtar, Zaki and Zuberi, Zia and Gallagher, Mark M},
title = {Feasibility of a Cryo‐Anchoring Catheter for Linear Ablation},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2025},
doi = {10.1111/jce.70131}
}
EUVIMED