An Ex Vivo Evaluation of Air Intrusion Into Pulsed Field Ablation Sheaths During Ablation and Mapping Catheter Insertion
Journal of Cardiovascular Electrophysiology · 2025-12 · 36 (12) · 3231-3237
DOI: 10.1111/jce.70129 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 02.10.2025. Verfasst von Chad Gier und 15 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Northwestern University, Chicago, Illinois, USA. Hauptthemen laut MEDLINE: Catheter Ablation, Cardiac Catheters, Air, Electrophysiologic Techniques, Cardiac. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The authors received no specific funding for this study..
Abstract (Originalsprache)
Introduction
Pulsed field ablation (PFA) is the newest ablation technology, and currently, no data exist on the amount of air intrusion into new, large bore PFA sheaths during ablation or mapping catheter insertion.
Methods
An ex vivo study was performed using various combinations of commercially available PFA ablation catheters and sheaths. Common mapping catheters and a non‐PFA steerable sheath were also evaluated as a reference. The siphon principle was used to create negative pressure to simulate left atrial pressure during spontaneous inspiration. Ablation and mapping catheters were advanced to the end of the sheaths under negative pressure and then removed. Air was withdrawn from the sheaths and was measured in milliliters (mL).
Results
A total of 55 trials were performed. The average volume of air intrusion with all sheath/catheter combinations was 9.6 + 5.2 mL. The 13 Fr (inner diameter) Faradrive sheath (Boston Scientific Inc.) entrained significantly more air (16.5 + 4.1 mL) compared with the 12 Fr FlexCath Contour sheath (Medtronic Inc.) (6.1 + 2.7 mL, p < 0.01), 13 Fr Agilis NxT sheath (Abbott Inc.) (8.7 + 1.8 mL, p < 0.01), and Vizigo sheath (Johnson & Johnson MedTech Inc.) (5.8 + 2.1 mL, p < 0.01), regardless of the catheter used. There was significantly higher volume of air intrusion with the Farawave ablation catheter through the Faradrive sheath (13.6 + 2.0 mL) than through the 13 Fr Agilis (9.4 + 2.1 mL, p = 0.03) or the PulseSelect through FlexCath Contour sheath (4.0 + 2.7 mL, p < 0.01). Mapping catheters entrained significantly more air than ablation catheters in both the Faradrive (18.0 ± 4.1 mL vs. 13.6 ± 4.1 mL, p = 0.04) and the FlexCath Contour (7.1 ± 2.2 mL vs. 4.0 ± 2.7 mL, p = 0.03).
Conclusion
Using a model that simulates left atrial PFA in spontaneously breathing patients, a large volume of air intrusion was observed during insertion and removal of ablation and mapping catheters into new, large bore PFA sheaths. New sheath designs are needed to minimize air intrusion during catheter exchanges to avoid air embolism when performing PFA using left atrial delivery sheaths.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Cardiology, Northwestern University, Chicago, Illinois, USA.
- University of Illinois College of Medicine, Chicago, Illinois, USA.
- Department of Internal Medicine, Northwestern University, Chicago, Illinois, USA.
ORCID-Profile: Chad Gier, Kaustubha D Patil, Alexandru Chicos, Albert C Lin, Rod S Passman, Nishant Verma
Förderung
- The authors received no specific funding for this study.
Zitieren
APA
Gier, C., Simon, E., Ahmed, A., Peigh, G., Patel, S., Baman, J., Kalluri, A., Culler, K., Patil, K. D., Pfenniger, A., Chicos, A., Kim, S. S., Lin, A. C., Passman, R. S., Knight, B. P., & Verma, N. (2025). An Ex Vivo Evaluation of Air Intrusion Into Pulsed Field Ablation Sheaths During Ablation and Mapping Catheter Insertion. Journal of Cardiovascular Electrophysiology, 36(12), 3231-3237. https://doi.org/10.1111/jce.70129
Vancouver
Gier C, Simon E, Ahmed A, Peigh G, Patel S, Baman J, et al. An Ex Vivo Evaluation of Air Intrusion Into Pulsed Field Ablation Sheaths During Ablation and Mapping Catheter Insertion. Journal of Cardiovascular Electrophysiology. 2025;36(12):3231-3237. doi:10.1111/jce.70129
BibTeX
@article{Gier2025,
author = {Gier, Chad and Simon, Erik and Ahmed, Aamir and Peigh, Graham and Patel, Shivam and Baman, Jayson and Kalluri, Aravind and Culler, Kasen and Patil, Kaustubha D and Pfenniger, Anna and Chicos, Alexandru and Kim, Susan S and Lin, Albert C and Passman, Rod S and Knight, Bradley P and Verma, Nishant},
title = {An Ex Vivo Evaluation of Air Intrusion Into Pulsed Field Ablation Sheaths During Ablation and Mapping Catheter Insertion},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2025},
volume = {36},
number = {12},
pages = {3231--3237},
doi = {10.1111/jce.70129}
}
EUVIMED