Predictive Role of Intraoperative Impedance in Midterm Pacing Threshold Elevation: Insights From Aveir VR Leadless Pacemaker Implantations

Takehiro Nomura iD, Kennosuke Yamashita iD, Michio Nagashima, Rei Kuji, Yosuke Mizuno, Daiki Kumazawa, Kosuke Onodera, Kenji Ando

Journal of Cardiovascular Electrophysiology · 2025

DOI: 10.1111/jce.16639 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2025. Verfasst von Takehiro Nomura und 7 weiteren Autor:innen. Beteiligte Einrichtung: Heart Rhythm Center, Department of Cardiovascular Medicine, Sendai Kosei Hospital, Sendai, Miyagi, Japan. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Bradycardia, Cardiac Pacing, Artificial, Heart Rate. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 19 Literaturangaben. Gefördert von: The authors received no specific funding for this work..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Leadless pacemakers (LPs) are a valuable treatment for bradycardia, with the Aveir offering advanced features, including a protective sleeve and active fixation. This study investigated the intraoperative factors during Aveir VR implantations that predict midterm pacing capture threshold (PCT) elevations and proposed strategies for optimizing the outcomes. Methods and Results This retrospective study included 67 patients who underwent Aveir VR implantations in two Japanese facilities between March and June 2023, with a 360‐day follow‐up. The impedance and PCT were measured three times during the procedure (premapping, during the tether mode, and at the end of the procedure). A PCT elevation was defined as that > 1.5 V/0.4 ms during follow‐up. PCT elevations were observed in 7 patients (10.4%), all with sinus node dysfunction. The impedance during the tether mode (Imp te ) and end of the procedure (Imp end ) was significantly lower in patients with a PCT elevation ( p < 0.001). The receiver operating characteristic curve analysis demonstrated that the Imp te (AUC: 0.868) and Imp end (AUC: 0.857) effectively predicted PCT elevations, with a cutoff of > 380 Ω associated with a lower incidence of PCT elevations. For the Imp te , the sensitivity was 85.7% and specificity 90.0%. Moreover, a high impedance during premapping (AUC: 0.892) predicted an Imp te > 380 Ω, and when the premapping impedance was > 430 Ω, all cases had an Imp te > 380 Ω. The median number of days until the PCT elevation was 40 (range: 1–321 days), and in 3 out of 7 patients, PCT elevations were observed more than 270 days after the implantation.

Conclusion

The higher impedance during the implantation was a strong predictor of a midterm favorable PCT. Those findings highlighted the potential for impedance‐guided strategies to optimize implantations and output settings. The occurrence of PCT elevations beyond 270 days underscored the need for long‐term monitoring and individualized follow‐up strategies. Further studies are needed to validate these findings and explore their clinical impact.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pacemaker, ArtificialBradycardiaCardiac Pacing, ArtificialHeart RateHumansMaleFemaleRetrospective StudiesAgedTreatment OutcomePredictive Value of TestsTime FactorsMiddle AgedAction PotentialsAged, 80 and overElectric ImpedanceJapanRisk FactorsEquipment Design

Stichwörter der Autor:innen

active‐fixationimpedanceleadless pacemakerpacing threshold

Autor:innen und Institutionen

  • Heart Rhythm Center, Department of Cardiovascular Medicine, Sendai Kosei Hospital, Sendai, Miyagi, Japan.
  • Department of Cardiovascular Medicine, School of Medicine, Dokkyo Medical University, Shimotsuga-gun, Tochigi, Japan.
  • Department of Cardiology, Kokura Memorial Hospital, Kitakyushu, Fukuoka, Japan.

ORCID-Profile: Takehiro Nomura, Kennosuke Yamashita

Förderung

  • The authors received no specific funding for this work.

Zitieren

APA

Nomura, T., Yamashita, K., Nagashima, M., Kuji, R., Mizuno, Y., Kumazawa, D., Onodera, K., & Ando, K. (2025). Predictive Role of Intraoperative Impedance in Midterm Pacing Threshold Elevation: Insights From Aveir VR Leadless Pacemaker Implantations. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16639

Vancouver

Nomura T, Yamashita K, Nagashima M, Kuji R, Mizuno Y, Kumazawa D, et al. Predictive Role of Intraoperative Impedance in Midterm Pacing Threshold Elevation: Insights From Aveir VR Leadless Pacemaker Implantations. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.16639

BibTeX
@article{Nomura2025,
  author = {Nomura, Takehiro and Yamashita, Kennosuke and Nagashima, Michio and Kuji, Rei and Mizuno, Yosuke and Kumazawa, Daiki and Onodera, Kosuke and Ando, Kenji},
  title = {Predictive Role of Intraoperative Impedance in Midterm Pacing Threshold Elevation: Insights From Aveir VR Leadless Pacemaker Implantations},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2025},
  doi = {10.1111/jce.16639}
}
Kostenlos starten →