Implantation Success, Electrical Performance, and Safety of an Active Fixation Stylet‐Driven Lead for LBBAP in Clinical Practice: A Multicenter Experience

Amato Santoro iD, Matteo Ziacchi, Raimondo Calvanese, Antonio Rapacciuolo, Mario Volpicelli, Luca Poggio, Pasquale Nocerino, Davide Castagno, Valter Bianchi, Paolo Liberato Magliano iD, Gabriele Dell'Era iD, Cosimo Mandurino und 12 weitere

Journal of Cardiovascular Electrophysiology · 2025

DOI: 10.1111/jce.70237 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2025. Verfasst von Amato Santoro und 23 weiteren Autor:innen. Beteiligte Einrichtung: Azienda Ospedaliera Universitaria Senese, Siena, Italy. Hauptthemen laut MEDLINE: Cardiac Pacing, Artificial, Heart Rate, Bundle of His, Pacemaker, Artificial. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Left bundle branch area pacing (LBBAP) has emerged as a promising physiological pacing strategy. However, its feasibility and safety using conventional stylet‐driven leads (SDL) such as INGEVITY+ remain under investigation.

Objective

To evaluate the clinical performance, safety, and follow‐up outcomes of LBBAP using the INGEVITY+ lead in a real‐world electrophysiology setting.

Methods

From 2021 to 2024, 207 consecutive patients underwent LBBAP implant attempts using the INGEVITY+ lead across 19 centers. Of these, 146 had bradycardia indications, and 61 had heart failure indications. A control group of 200 patients received standard right ventricular (RV) pacing with the same lead. Procedural outcomes, electrical parameters, complications, and mid‐term lead performance were assessed.

Results

LBBAP was successfully achieved in 201 (97%) cases. Procedural and fluoroscopy times were slightly longer for LBBAP than for RV pacing. Acute complications included one helix fracture, one atrioventricular block, and one septal perforation. During a median follow‐up of 9 months, no lead fractures were reported; lead dislodgment occurred in three cases. Kaplan–Meier analysis showed no significant difference in time to first lead‐related complication between the LBBAP and RV pacing groups (hazard ratio: 4.69, 95%CI: 0.63–34.90, p = 0.139). Electrical performance remained stable, with 98% of retained leads maintaining capture thresholds ≤ 2 V and 92% with sensed amplitudes ≥ 5 mV.

Conclusions

LBBAP using the INGEVITY + SDL is feasible and safe in clinical practice, with excellent implant success and stable mid‐term electrical performance. These findings support its use as a viable conduction system pacing strategy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiac Pacing, ArtificialHeart RateBundle of HisPacemaker, ArtificialBradycardiaHeart FailureBundle-Branch BlockHumansMaleFemaleAgedTreatment OutcomeMiddle AgedTime FactorsAction PotentialsRetrospective StudiesRisk FactorsEquipment DesignFeasibility StudiesAged, 80 and overRecovery of Function

Stichwörter der Autor:innen

conduction system pacingleft bundle branch area pacingpacemaker implantationstylet‐driven lead

Autor:innen und Institutionen

  • Azienda Ospedaliera Universitaria Senese, Siena, Italy.
  • S. Orsola-Malpighi University Hospital, University of Bologna, Bologna, Italy.
  • Ospedale del Mare, Naples, Italy.
  • Federico II University of Naples, Naples, Italy.
  • Ospedale Santa Maria della Pietà, Nola, Italy.
  • Ospedale Maggiore di Lodi, Lodi, Italy.

ORCID-Profile: Amato Santoro, Paolo Liberato Magliano, Gabriele Dell'Era, Carmelo La Greca

Zitieren

APA

Santoro, A., Ziacchi, M., Calvanese, R., Rapacciuolo, A., Volpicelli, M., Poggio, L., Nocerino, P., Castagno, D., Bianchi, V., Magliano, P. L., Dell'Era, G., Mandurino, C., Mirizzi, G., Savarese, G., Marinaccio, L., La Greca, C., Covino, G., Zingarini, G., Varalda, M., …, & Bertini, M. (2025). Implantation Success, Electrical Performance, and Safety of an Active Fixation Stylet‐Driven Lead for LBBAP in Clinical Practice: A Multicenter Experience. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70237

Vancouver

Santoro A, Ziacchi M, Calvanese R, Rapacciuolo A, Volpicelli M, Poggio L, et al. Implantation Success, Electrical Performance, and Safety of an Active Fixation Stylet‐Driven Lead for LBBAP in Clinical Practice: A Multicenter Experience. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.70237

BibTeX
@article{Santoro2025,
  author = {Santoro, Amato and Ziacchi, Matteo and Calvanese, Raimondo and Rapacciuolo, Antonio and Volpicelli, Mario and Poggio, Luca and Nocerino, Pasquale and Castagno, Davide and Bianchi, Valter and Magliano, Paolo Liberato and Dell'Era, Gabriele and Mandurino, Cosimo and Mirizzi, Gianluca and Savarese, Gianluca and Marinaccio, Leonardo and La Greca, Carmelo and Covino, Gregorio and Zingarini, Gianluca and Varalda, Marco and Vitali, Francesco and Baiocchi, Claudia and Campari, Monica and Valsecchi, Sergio and Bertini, Matteo},
  title = {Implantation Success, Electrical Performance, and Safety of an Active Fixation Stylet‐Driven Lead for LBBAP in Clinical Practice: A Multicenter Experience},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2025},
  doi = {10.1111/jce.70237}
}
Kostenlos starten →