Catheter–tissue contact optimizes pulsed electric field ablation with a large area focal catheter
Journal of Cardiovascular Electrophysiology · 2024
DOI: 10.1111/jce.16208 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2024. Verfasst von Shephal K Doshi und 4 weiteren Autor:innen. Beteiligte Einrichtung: Cardiology Electrophysiology, Pacific Heart Institute, Santa Monica, California, USA. Hauptthemen laut MEDLINE: Catheter Ablation, Atrial Fibrillation. Die Arbeit wurde bisher 25-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Cardiofocus.
Abstract (Originalsprache)
Introduction
Pulsed electric field (PEF) ablation relies on the intersection of a critical voltage gradient with tissue to cause cell death. Field‐based lesion formation with PEF technologies may still depend on catheter–tissue contact (CTC). The purpose of this study was to assess the impact of CTC on PEF lesion formation with an investigational large area focal (LAF) catheter in a preclinical model.
Methods
PEF ablation via a 10‐spline LAF catheter was used to create discrete right ventricle (RV) lesions and atrial lesion sets in 10 swine (eight acute, two chronic). Local impedance (LI) was used to assess CTC. Lesions were assigned to three cohorts using LI above baseline: no tissue contact (NTC: ≤∆10 Ω, close proximity to tissue), low tissue contact (LTC: ∆11–29 Ω), and high tissue contact (HTC: ≥∆30 Ω). Acute animals were infused with triphenyl tetrazolium chloride (TTC) and killed ≥2 h post‐treatment. Chronic animals were remapped 30 days post‐index procedure and stained with infused TTC.
Results
Mean (± SD) RV treatment sizes between LTC ( n = 14) and HTC ( n = 17) lesions were not significantly different (depth: 5.65 ± 1.96 vs. 5.68 ± 2.05 mm, p = .999; width: 15.68 ± 5.22 vs. 16.98 ± 4.45 mm, p = .737), while mean treatment size for NTC lesions ( n = 6) was significantly smaller (1.67 ± 1.16 mm depth, 5.97 ± 4.48 mm width, p < .05). For atrial lesion sets, acute and chronic conduction block were achieved with both LTC ( N = 7) and HTC ( N = 6), and NTC resulted in gaps.
Conclusions
PEF ablation with a specialized LAF catheter in a swine model is dependent on CTC. LI as an indicator of CTC may aid in the creation of consistent transmural lesions in PEF ablation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiology Electrophysiology, Pacific Heart Institute, Santa Monica, California, USA.
- Electrophysiology, Galvanize Therapeutics, Redwood City, California, USA.
- CardioFocus, Marlborough, Massachusetts, USA.
- Department for Cardiovascular Diseases, Klinički Bolnički Centar (KBC) Split, Split, Croatia.
ORCID-Profile: Mary Clare Flaherty
Förderung
- Cardiofocus
Zitieren
APA
Doshi, S. K., Flaherty, M. C., Laughner, J., Quan, M., & Anic, A. (2024). Catheter–tissue contact optimizes pulsed electric field ablation with a large area focal catheter. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16208
Vancouver
Doshi SK, Flaherty MC, Laughner J, Quan M, Anic A. Catheter–tissue contact optimizes pulsed electric field ablation with a large area focal catheter. Journal of Cardiovascular Electrophysiology. 2024. doi:10.1111/jce.16208
BibTeX
@article{Doshi2024,
author = {Doshi, Shephal K and Flaherty, Mary Clare and Laughner, Jacob and Quan, Melinda and Anic, Ante},
title = {Catheter–tissue contact optimizes pulsed electric field ablation with a large area focal catheter},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2024},
doi = {10.1111/jce.16208}
}
EUVIMED