Utility and Limitations of Ablation Index for Guiding Therapy in Ventricular Myocardium
JACC: Clinical Electrophysiology · 2023
DOI: 10.1016/j.jacep.2023.03.020 ↗Kurz gefasst
Fachartikel in der Zeitschrift JACC: Clinical Electrophysiology, 2023. Verfasst von Arwa Younis und 5 weiteren Autor:innen. Beteiligte Einrichtung: Cardiac Electrophysiology Section, Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio, USA. Hauptthemen laut MEDLINE: Catheter Ablation. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 15 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: Ablation index (AI) is used for guiding therapy during pulmonary vein isolation. However, its potential utility in ventricular myocardium is unknown. OBJECTIVES: This study sought to examine the correlation between AI and lesion dimensions in healthy and infarcted ventricles. METHODS: In ex vivo experiments using healthy swine ventricles, the correlation between AI (400-1,200) and lesion dimensions was examined at fixed power (30 W) and contact force (CF) (15 g). To examine the accuracy of AI in predicting lesion dimensions created by different combinations of ablation parameters, applications with a similar prespecified AI value created using different power (30 vs 40 W), CF (15 vs 25 g) or impedance (130-170 Ω) were created. In in vivo experiments, the correlation between AI and lesion dimensions was examined in healthy and infarcted myocardium. RESULTS: Ex vivo experiments (247 lesions, 36 hearts) showed good correlation between AI and lesion depth (R = 0.93; P < 0.001). However, in vivo experiments (9 healthy swine and 10 infarcted swine) showed moderate correlation in healthy myocardium (R = 0.64; P < 0.01) and poor correlation in infarcted myocardium (R = 0.23; P = 0.61). AI values achieved using different combinations of power, CF, and baseline impedance resulted in different lesion depths: Ablation at 30 W produced deeper lesions compared with 40 W, ablation with CF of 15 g produced deeper lesions compared with CF of 25 g, and ablation at lower impedance produced larger lesions at similar prespecified AI values (P < 0.01 for all). CONCLUSIONS: AI has limited value for guiding ablation in ventricular myocardium, particularly scar. This may be related to small proportional significance of application duration and complex tissue architecture.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiac Electrophysiology Section, Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio, USA.
- Cardiac Electrophysiology Section, Division of Cardiovascular Medicine, Shamir Medical Center, Be'er Yaakov, Israel; Sackler School of Medicine, Tel Aviv University, Ramat Aviv, Tel Aviv, Israel.
- Cardiac Electrophysiology Section, Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio, USA; Cardiac Electrophysiology Section, Division of Cardiovascular Medicine, Shamir Medical Center, Be'er Yaak…
Zitieren
APA
Younis, A., Zilberman, I., Yavin, H., Higuchi, K., Barkagan, M., & Anter, E. (2023). Utility and Limitations of Ablation Index for Guiding Therapy in Ventricular Myocardium. JACC: Clinical Electrophysiology. https://doi.org/10.1016/j.jacep.2023.03.020
Vancouver
Younis A, Zilberman I, Yavin H, Higuchi K, Barkagan M, Anter E. Utility and Limitations of Ablation Index for Guiding Therapy in Ventricular Myocardium. JACC: Clinical Electrophysiology. 2023. doi:10.1016/j.jacep.2023.03.020
BibTeX
@article{Younis2023,
author = {Younis, Arwa and Zilberman, Israel and Yavin, Hagai and Higuchi, Koji and Barkagan, Michael and Anter, Elad},
title = {Utility and Limitations of Ablation Index for Guiding Therapy in Ventricular Myocardium},
journal = {JACC: Clinical Electrophysiology},
year = {2023},
doi = {10.1016/j.jacep.2023.03.020}
}
EUVIMED